List of Error Codes
Error code
Cause
How to correct
Er1
In ation error
When the pressure does not exceed 15 
mmHg after operating a pump for 15 seconds 
from the start of in ation
When the in ation does not reach the set 
pressure of the arm cuff within the speci ed 
time after starting the in ation
If the connector of rubber tube connecting the 
arm cuff and the main unit is loose, connect them 
correctly, then measure.
If the rubber tube connecting the arm cuff and the 
main unit is bent, stretch it, then measure.
If the arm cuff is wrapped loosely, wrap it correctly, 
then measure.
Er2
De ation error
When the de ation speed is too fast during 
the measurement
When the de ation speed is too slow during 
the measurement
When the measurement does not end 
within the speci ed time after starting the 
measurement
When the pressure is applied more than 5 
minutes
Er3
Overpressure error
The arm cuff pressure exceeded 299 mmHg.
If the rubber tube connecting the arm cuff and the 
main unit is bent, stretch it, then measure.
Er4
Insuf cient in ation error 
Blood pressure could not be measured due 
to insuf cient in ation after the measurement 
ends.
If the measurement is made by setting P-SET to 
“AUTO”, ask the patient not move his/her arm or 
body during the in ation.
When setting the P-SET to “AUTO”, turn the Volume 
counterclockwise as far as it goes until you can hear 
the click sound.
Set the P-SET to 30 to 40 mmHg higher than the 
systolic pressure of the patient, then measure.
Er5
Indeterminable blood pressure error
Blood pressure could not be measured even 
when the arm cuff pressure reached the 
speci ed pressure.
If the arm cuff is wrapped loosely, wrap it correctly, 
then measure.
Er6
Pulse wave small error 
Pulse wave was too small.
If the arm cuff is wrapped loosely, wrap it correctly, 
then measure.
Do not wrap the arm cuff over a thick cloth or by 
rolling up the jacket or shirt.
Er7
Measurement error
Relationship between systolic and diastolic 
pressures was abnormal.
Check if the patient has an arrhythmia.
Ask the patient not to move his/her arm or body 
during the measurement.

Er8
Pulse rate error
Pule rate did not stay within the range of 30 
to 199 beats/min.
Er9
Device error
There is a problem with the main unit of the 
blood pressure monitor.
Contact the dealer where you purchased the unit or 
the nearest OMRON dealer.