Hanna Instruments hi 99300 User Manual

Page of 2
Instruction Manual
HI 99300 • HI 99301
Portable Instruments
for EC/TDS/°C/°F measurements
with advanced features
w w w . h a n n a i n s t . c o m
    IST99300R1       01/05
SPECIFICATIONS
HI 99300
HI 99301
Range
EC
0 to 3999 µS/cm
0.00 to 20.00 mS/cm
TDS
0 to 2000 ppm (mg/l)
0.00 to 10.00 ppt (g/l)
Temperature                  0.0 to 60.0°C / 32.0 to 140.0°F
Resolution
EC
1 µS/cm
0.01 mS/cm
TDS
1 ppm (mg/l)
0.01 ppt (g/l)
Temperature
0.1°C / 0.1°F
Accuracy
 (@20°C/68°F)
±2% F.S. (EC/TDS)
±0.5°C / ±1°F
Typical EMC Deviation    ±2% F.S. (EC/TDS) / ±0.5°C / ±1°F
Temperature
Automatic, from 0 to 60°C
Compensation
β
 adjustable from 0.0 to 2.4%/°C (BETA)
Calibration
Automatic, single point
EC/TDS Factor
Selectable from 0.45 to 1.00 (CONV)
Probe  (included)
HI 76306 with built-in temperature sensor,
DIN connector and 1 m cable
Environment
0 to 50°C (32 to 122°F); RH 100%
Power Supply
4 alkaline batteries, 1.5V AAA with BEPS;
500 hours of continuous use
Dimensions
143x80x38 mm (5.6x3.2x1.5’’)
Weight
320 g (11.3 oz.)
HI 99300 and HI 99301 are in compliance with the 
 directives.
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in
workmanship and materials when used for their intended purpose and main-
tained according to instructions. The probes are warranted for six months.
This warranty is limited to repair or replacement free of chargeDamage due to
accidents, misuse, tampering or lack of prescribed maintenance are not
covered. If service is required, contact the dealer from whom you purchased
the instrument. If under warranty, report the model number, date of purchase,
serial number and the nature of the problem. If the repair is not covered by the
warranty, you will be notified of the charges incurred. If the instrument is to be
returned to Hanna Instruments, first obtain a Returned Goods Authorization
Number from the Customer Service department and then send it with shipment
costs prepaid. When shipping any instrument, make sure it is properly
packaged for complete protection.
DISPLAY FUNCTIONAL DESCRIPTION
HI 99300 and HI 99301 come supplied complete with HI 76306 probe,
4 alkaline batteries (1.5V AAA), instructions and rugged carrying case.