Hanna Instruments hi 9043 User Manual

Page of 2
w w w . h a n n a i n s t . c o m
Hanna Instruments reserves the right to modify the design, construction and
appearance of its products without advance notice.
WARRANTY
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for correct opera-
tion. Please read it carefully before using the meter. If you
need additional technical information, do not hesitate to e-
mail us at tech@hannainst.com.
These instruments are in compliance with the       direc-
tives.
PRELIMINARY EXAMINATION
GENERAL DESCRIPTION
FUNCTIONAL DESCRIPTION
These Instruments are in
Compliance with CE
Directives
SPECIFICATIONS
Instruction Manual
HI 9043 - HI 9044
Portable Microprocessor
K-Type Thermocouple
Thermometers
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipping. If there is any noticeable damage,
notify your Dealer or the nearest Hanna Office.
Note: Save all packing materials until you are sure that the
instrument functions correctly. Any defective item must
be returned in the original packaging together with
the supplied accessories.
Range (*)
Resolution
Accuracy
(@20°C/68°F)
Typical EMC
Deviation
Display
Battery Type
Life
Probe
Environment
Dimensions
Weight
1. Liquid Crystal Display
2. Measuring unit selection key
3. HI/LO values reset key
4. HOLD key
5. ON/OFF key
HI 9043 and HI 9044 are  microprocessor-based thermometers
which allow temperature measurements using K-type thermo-
couple probes. The non-linearity of the temperature probe is
linearized by the microprocessor.
Standard features include dual-level LCD for simultaneously
displaying of the highest and lowest measured temperatures,
display hold, selectable measurement unit (°C or °F) and
resolution which  automatically switches from 0.1° to 1° for
temperatures above 200°C (400°F).
The meters are also provided with low battery detection and
BEPS (Battery Error Preventing System), which turns the
unit off when the battery is discharged avoiding erroneous
readings caused by low battery level.
HI 9043 comes complete with a 9V battery and instruction
manual.
HI 9044 comes complete with HI 766E2 fixed probe, a 9V
battery and instruction manual.
1) Current temperature value
2) HOLD indicator
3) Low Battery indicator
4) Minimum temperature value
5) Maximum temperature value
6) K-type probe indicator
7) Measuring unit, °C or °F
FACTORY RECALIBRATION
All Hanna thermometers have been accurately pre-cali-
brated at the factory. It is generally recommended to have
all thermometers recalibrated at least once a year.
For an accurate annual recalibration, contact your dealer or
the nearest Hanna Service Center.
All Hanna Instruments meters are warranted for two years against
defects in workmanship and materials when used for their intended pur-
pose and maintained according to instructions. Sensors and probes are
warranted for a period of six months.
This warranty is limited to repair or replacement free of charge.
Damages due to accidents, misuse, tampering or lack of prescribed main-
tenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase,
serial number and the nature of the failure.
First obtain a Returned Goods Authorization number from the Customer
Service department, then return the instrument with the Authorization #
included along with shipment costs prepaid. If the repair is not covered by
the warranty, you will be notified of the charges.
When shipping any instrument, make sure it is properly packaged for
complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments
Inc., Woonsocket, Rhode Island, 02895 , USA.
-50.0 to 199.9°C / 200 to 1350°C
-58.0 to 399.9°F / 400 to 2462°F
0.1°C (up to 199.9°C) / 1°C (outside)
0.1°F (up to 399.9°F) / 1°F (outside)
±0.2% F.S.
excluding probe error
±3 °C and ±6 °F
with HI 766 K-Thermocouple probe
Dual-line LCD
1 x 9V (IEC 6LR61) battery
approx. 500 hours of continuous use
HI 766 K-Thermocouple series
• optional for HI 9043
• HI 766E2 fixed for HI 9044
-10 to 50ºC (14 to 122ºF);
max. 95% RH non-condensing
180 x 83 x 40 mm (7.1x3.3x1.6")
226 g (8 oz.) meter only
(*) Range may be limited by probe. The range of HI 9044
is limited to 900°C/1650°F by the fixed probe.