Hanna Instruments c 125 User Manual

Page of 16
4
Remove the instrument from the packing material and examine it to make sure that
no damage has occurred during shipping. If there is any damage, notify your Dealer.
C125 is supplied complete with:
• Glass Cuvette & Cap (2 pcs.)
• Calibration solutions HI93703-0, HI93703-10 and HI93703-5
• Tissue for wiping cuvettes (HI731318)
• 1.5V AA alkaline batteries (4 pcs.)
• Instruction manual
• Rugged carrying case
Note: Save all packing material until you are sure that the meter functions
correctly. All defective items must be returned in its original packaging
with the supplied accessories.
PRELIMINARY EXAMINATION
C125 has been designed to perform measurements according to the ISO 7027
International Standard.
The instrument functions by passing a beam of infrared light through a vial
containing the sample being measured.
The light source is a High Emission Infrared LED with a wavelength peaking at 890
nm, ensuring that the interference caused by colored samples is minimum.
A sensor, positioned at 90° with respect to the direction of light, detects the amount
of light scattered by the undissolved particles present in the sample.
The microprocessor converts such readings into FTU (=FNU) values.