Hanna Instruments c 99 User Manual

Page of 148
101
• Press READ DIRECT and "SIP" will appear during measurement.
• The instrument directly displays concentration in mg/L of nickel on
the Liquid Crystal Display.
Note: a temperature above 30°C may cause turbidity. In this
case, before zeroing and taking readings, add 2-3 drops of
Dispersing Agent (HI 93703-51) to each cuvet and swirl until
complete dissolution of the turbidity.
INTERFERENCES
Interference may be caused by:
Co
2+
 must not be present
Fe
2+
 must not be present
Al
3+
 above 32 mg/L
Ca
2+
 above 1000 mg/L (as CaCO
3
)
Cd
2+
 above 20 mg/L
Cl
 above 8000 mg/L
Cr
3+
 above 20 mg/L
Cr
6+
 above 40 mg/L
Cu
2+
 above 15 mg/L
F
 above 20 mg/L
Fe
3+
 above 10 mg/L
K
+
 above 500 mg/L
Mg
2+
 above 400 mg/L
Mn
2+
 above 25 mg/L
Mo
6+
 above 60 mg/L
Na
+
 above 5000 mg/L
Pb
2+
 above 20 mg/L
Zn
2+
 above 30 mg/L
Nickel LR