Hanna Instruments hi 983315-0 User Manual

Page of 2
Instruction Manual
HI 983315-0
HI 983315-1
HI 98331500
Panel-Mounted
TDS Indicators &
Controllers
Manufacturers since 1978
These Instruments are in
Compliance with the CE Directives
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
WARRANTY
WARRANTY
WARRANTY
WARRANTY
WARRANTY
These meters are warranted for two years against defects in work-
manship and materials when used for their intended purpose and
maintained according to instructions. The probe is warranted for a
period of six months. This warranty is limited to repair or replace-
ment free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-
chase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service de-
partment and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protec-
tion.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instru-
ments Inc., Woonsocket, Rhode Island, 02895 , USA.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the   directives.
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
HI983315-0, HI983315-1 and HI98331500 are conductiv-
ity  indicators and controllers with a relay output designed
for simplicity of use in a wide range of industrial applica-
tions.
HI98331500 is provided with a DC/DC converter which
isolates the measure system from the external power supply.
They are panel-mounted with membrane keypads on the
front panel and an easy-to-read LCD display.
All the meters automatically compensate for temperature
variations. The probe is easy to clean and requires little
maintenance.
Measurements are highly accurate and the meters can be
calibrated at one point.
Power supply, wiring and selection are made via the plug-
in terminal blocks on the rear panel.
LED indicators on the front panel identify whether the
controller is in set or reading mode; alarm led lights up
when measurements are higher than selected setpoint value.
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range
0.0 to 199.9 ppm
Resolution
 0.1 ppm
Accuracy (@ 20°C)
±2% f.s.
Setpoint
Adjustable through multiturn trimmer
Alarm Condition LED ON and alarm contact closed when
measured TDS value is higher than setpoint
Alarm  Output
2-contact relay, no fuse protected.
5A, 240 VAC, 30 VDC
Probe
HI 7634-00 EC/TDS probe
Temp. Compensation Automatic from 5 to 50°C
(41 to 122°F) with ß=2%
TDS Conversion Factor
  0.5
Calibration
Manual, at one point through trimmer
Calibration Solution
HI 7033 (84 µS/cm)
Power supply:
 External
HI983315-0
12  VDC
HI98331500
12VDC  isolated inside with DC/DC converter
HI983315-1
 115/230 VAC; 50/60Hz
Dimensions
 79 x 49 x 95 mm (3.1 x 1.9 x 3.7”)
Weight:
HI983315-0, HI98331500
140 g (4.96 oz)
HI983315-1
 250 g (8.85 oz)
CE DECLARATION
CE DECLARATION
CE DECLARATION
CE DECLARATION
CE DECLARATION
OF CONFORMITY
OF CONFORMITY
OF CONFORMITY
OF CONFORMITY
OF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely suitable for
the environment in which they are used. Operation of these instruments in
residential areas could cause unacceptable interferences to radio and TV
equipment. The metal band at the end of the probe is sensitive to electro-
static discharges. Avoid touching this metal band at all times. During op-
eration, ESD wrist straps should be worn to avoid possible damage to the
probe by electrostatic discharges. Any variation introduced by the user to
the supplied equipment may degrade the instrument’s EMC performance.
To avoid electrical shock, do not use these instruments when voltage at the
measurement surface exceeds 24 VAC or 60 VDC. Use plastic beakers to
minimize any EMC interferences.
To avoid damage or burns, do not perform any measurement in microwave
ovens.
ISTR983315R5      12/2001
h t t p : / / w w w . h a n n a i n s t . c o m
Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
The meters are supplied with:
• Instruction manual;
• Mounting brackets.
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.