Hanna Instruments hi 991404 User Manual

Page of 2
w w w . h a n n a i n s t . c o m
Hanna Instruments reserves the right to modify the design, construction and
appearance of its products without advance notice.
WARRANTY
HI 991404 and HI 991405 are warranted for two years against
defects in workmanship and materials when used for their intended pur-
pose and maintained according to instructions. The probes are
warranted for a period of six months. This warranty is limited to
repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed mainte-
nance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase,
serial number and the nature of the failure. If the repair is not covered by
the warranty, you will be notified of the charges incurred. If the instrument
is to be returned to Hanna Instruments, first obtain a Returned Goods
Authorization Number from the Customer Service department and then
send it with shipment costs prepaid. When shipping any instrument, make
sure it is properly packaged for complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments
Inc., Woonsocket, Rhode Island, 02895 , USA.
Instruction Manual
(HI 991404 - HI 991405)
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for a correct
operation. Please read it carefully before using the meter.
If you need additional technical information, do not hesitate
to e-mail us at tech@hannainst.com.
These instruments are in compliance with the        directives.
Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
The meters are supplied with:
 pH 4.01 and 7.01 buffer solutions (20 mL each);
• HI 70031 1413 µS/cm calibration solution, for HI 991404;
• HI 70030 12.88 mS/cm calibration solution, for HI 991405;
• HI 1293D pH-electrode with differential input signal;
• Hanna 12 VDC power adapter;
• Instruction manual.
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATION
COMBO 
 
is a combined pH/temperature/EC/TDS
meter specially designed to meet the needs of growers in
greenhouses and hydroponic applications.
You can simply install the meter above the sample to be
tested for continuous measurement.
The  HI 1293D pH-electrode & the EC/TDS/Temperature
probe have been molded with pipe thread (1/2’’ NPT),
which allows the user to attach them to an in-line system.
Measurements are accurate and the meter can be calibrated
at one or two points for pH and at single point for EC/TDS.
You no longer need to worry about battery changes either:
the unit runs without interruption on 12 VDC power supply.
GENERAL DESCRIPTION
FUNCTIONAL DESCRIPTION
These Instruments are in
Compliance with CE
Directives
SPECIFICATIONS
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the
environment in which it is used. Operation of this instrument in residen-
tial areas could cause unacceptable interferences to radio and TV equip-
ment.
The glass bulb at the end of the electrode is sensitive to electrostatic dis-
charges. Avoid touching this glass bulb at all times. During operation, ESD
wrist straps should be worn to avoid possible damage to the electrode by
electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade
the instrument’s EMC performance.
To avoid electrical shock, do not use this instrument when voltages at the
measurement surface exceed 24 VAC or 60 VDC. To avoid damages or burns,
do not perform any measurement in microwave ovens.
ISTR991404R1    05/02
HI 991404
EC/TDS Range
0 to 3999 µS/cm - 0 to 2000 ppm
Resolution
1 µS/cm - 1 ppm
Accuracy 
(@20ºC/68ºF)
±2% f.s.
Typical EMC Deviation
±2.5 f.s.
HI 991405
EC/TDS Range 0.00 to 20.00 mS/cm - 0.00 to 10.00 ppt
Resolution
0.01 mS/cm - 0.01 ppt
Accuracy 
(@20ºC/68ºF)
±2% f.s.
Typical EMC deviation
±2.5% f.s.
Common specifications
pH Range
0.0 to 14.0 pH
Temperature Range
0.0 to 60ºC (32.0 to 140ºF)
Resolution
0.1 pH
0.1ºC or 0.1ºF
Accuracy 
(@20ºC/68ºF)
±0.1 pH
±0.5 ºC or ±1ºF
Typical EMC Deviation
±0.1 pH
±1 ºC or ±2 ºF
Temp.Compensation     pH
Automatic
                     EC/TDS     
β
=0.0 to 2.4%/ºC
TDS Conversion factor
0.45 to 1.00 (CONV)
pH electrode
HI 1293D (included)
pH Calibration
At 1 or 2 points with auto-buffer recognition
EC/TDS Calibration
Automatic at 1 point
Power supply
   12VDC power adapter (included)
Environment
0 to 50ºC (32 to 122ºF);
RH 95% non-condensing
Dimensions (meter only)
160 x 105 x 31 mm
(6.2 x 4.1 x 1.2’’)
Weight (meter only)
190 g (6.7 oz.)
1. Liquid Crystal Display
2. ON/OFF/MODE button
3. Power supply connector
4. 12 VDC power adapter
5. HI 1293D pH-electrode with differential input, pipe
thread 1/2’’ NPT
6. EC/TDS/Temperature probe, pipe thread 1/2’’ NPT
(works also as matching pin for pH-electrode)
7. DIN connector
8. SET/HOLD button