Hanna Instruments hi 98129 User Manual

Page of 2
w w w . h a n n a i n s t . c o m
Hanna Instruments reserves the right to modify the design, con-
struction and appearance of its products without advance notice.
WARRANTY
WARRANTY
WARRANTY
WARRANTY
WARRANTY
HI 98129 and HI 98130 are warranted for one
year against defects in workmanship and materials
when used for their intended purpose and maintained
according to instructions. The electrode is war-
ranted for a period of six months. This warranty is
limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack
of prescribed maintenance are not covered.
If service is required, contact the dealer from whom
you purchased the instrument. If under warranty,
report the model number, date of purchase, serial
number and the nature of the failure. If the repair is
not covered by the warranty, you will be notified of
the charges incurred. If the instrument is to be re-
turned to Hanna Instruments, first obtain a Returned
Goods Authorization Number from the Customer
Service department and then send it with shipment
costs prepaid. When shipping any instrument, make
sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the
enclosed warranty card within 14 days from the date
of purchase.
All rights are reserved. Reproduction in whole or in
part is prohibited without the written consent of the
copyright owner, Hanna Instruments Inc., Woonsocket,
Rhode Island, 02895 , USA.
Instruction Manual
HI 98129 - HI 98130
Waterproof pH, EC/TDS &
temperature meters
Dear Customer,
Thank you for choosing a Hanna product. This manual
will provide you with the necessary information for a
correct operation. Please read it carefully before
using the meter. If you need additional technical
information, do not hesitate to e-mail us at
tech@hannainst.com.
These instruments are in compliance with the
directives.
Remove the instrument from the packing material
and examine it carefully. If any damage has occurred
during shipment, immediately notify your Dealer or
the nearest Hanna Customer Service Center.
The meters are supplied with:
• 4 x 1.5V batteries
• HI 73127 pH electrode
• HI 73128 Tool to remove the pH electrode
Note: Conserve all packing material until the instru-
ment has been observed to function correctly.
Any defective item must be returned in its
original packing.
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
HI 98129 and HI 98130 are waterproof pH, EC/TDS
and temperature meters. The housing has been com-
pletely sealed against humidity and designed to float.
All pH and EC/TDS readings are automatically tem-
perature compensated (ATC), and temperature values
can be displayed in °C or °F units.
For EC/TDS readings, the EC/TDS conversion factor
(CONV) is selectable by the user, as well as the
temperature compensation coefficient 
β
 (BETA).
The meters can be calibrated at one or two points for
pH (with auto-buffer recognition and against five
memorized buffer values), and at one point for EC/
TDS.
Measurements are highly accurate with a unique
stability indicator right on the LCD.
A low battery symbol warns the user when the batter-
ies are to be replaced. In addition the Battery Error
Prevention System (BEPS) avoids erroneous reading
caused by low voltage level by turning the meter off.
The HI 73127 pH electrode, supplied with the meters,
is interchangeable and can be easily replaced by the
user.
The stainless steel encapsulated temperature sensor
facilitates faster and more accurate temperature mea-
surement and compensation.
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
                                                                                                                                                                                                                                                                
1. Battery compartment
2. Liquid Crystal Display (LCD)
3. Stability indicator
4. Low battery indicator
5. ON/OFF/MODE button
6. HI 73127 pH electrode
7. Temperature sensor
8. EC/TDS probe
9. SET/HOLD button
10. Secondary LCD
11. Primary LCD
This Instrument is in
Compliance with CE
Directives
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range Temperature:
0.0 to 60.0°C or 32.0 to 140.0°F
HI 98129
pH: 0.00 to 14.00
EC: 0 to 3999 µS/cm
TDS: 0 to 2000 ppm
HI 98130
pH: 0.00 to 14.00
EC: 0.00 to 20.00 mS/cm
TDS: 0.00 to 10.00 ppt
Resolution
0.1°C or 0.1°F
   HI 98129
0.01 pH; 1 µS/cm ; 1 ppm
   HI 98130
0.01 pH, 0.01 mS/cm ; 0.01 ppt
Accuracy 
(@20°C/68°F)
Temperature ±0.5°C or ±1°F
EC/TDS
±2% f.s.
pH
±0.01
Typical  EMC Deviation Temperature ±0.5°C or ±1°F
pH
±0.02 pH
EC/TDS
±2% f.s.
Temp.  Compensation       pH: Automatic
                               EC/TDS: with 
β
=0.0 to 2.4%/ºC
Environment
0 to 50°C (32 to 122°F); RH 100%
EC/TDS Conversion Factor 0.45 to 1.00 (CONV)
Calibration
pH: at 1 or 2 points with 2 sets of memo
      rized buffers (pH 4.01/7.01/10.01 or
      pH 4.01/6.86/9.18)
EC/TDS: automatic, at 1 point
EC/TDS Calibration Solutions
HI 98129
HI7031 (1413 µS/cm)
HI7032 (1382 ppm; CONV=0.5)
HI70442 (1500 ppm; CONV=0.7)
HI 98130
HI7030 (12.88 mS/cm)
HI70038 (6.44 ppt; CONV=0.5
                or 9.02 ppt; CONV=0.7)
Electrode
HI 73127     pH electrode (included
included
included
included
included)
Battery 
 Type/Life      4 x 1.5V with BEPS / typical 100 hours
Auto-off
          After 8 min.
Dimensions      163 x 40 x 26 mm (6.4 x 1.6 x 1.0”)
Weight
          85 g (3.0 oz)
Recommendations for Users
Before using this product, make sure that it is entirely suit-
able for the environment in which it is used. Operation of
this instrument in residential areas could cause unaccept-
able interferences to radio and TV equipment.
The glass bulb at the end of the electrode is sensitive to elec-
trostatic discharges. Avoid touching this glass bulb at all times.
During operation, ESD wrist straps should be worn to avoid
possible damage to the electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equip-
ment may degrade the instrument’s EMC performance.
To avoid electrical shock, do not use this instrument when
voltages at the measurement surface exceed 24 VAC or 60
VDC. To avoid damages or burns, do not perform any mea-
surement in microwave ovens.
ISTR98129R2    01/02