Hanna Instruments hi 98109 User Manual
Skincheck
Skincheck
Skincheck
Skincheck
Operating Instructions
• Switch the meter on
and remove the protec-
tive cap.
tive cap.
• Place the tip of the flat
electrode on the sur-
face to be measured
(skin or scalp).
face to be measured
(skin or scalp).
• For best results, wet
the surface with dis-
tilled water or a neutral disin-
fectant.
tilled water or a neutral disin-
fectant.
• The supplied electrode can
also measure pH in liquids.
Immerse the tip of the elec-
trode (the bottom 2cm/1") into
solution. Stir gently and wait
until the display stabilizes.
Immerse the tip of the elec-
trode (the bottom 2cm/1") into
solution. Stir gently and wait
until the display stabilizes.
• For best results, recalibrate pe-
riodically (see below).
• Do not be alarmed if white crys-
tals appear around the cap.
This is normal with pH elec-
trodes and they dissolve when
rinsed with water.
trodes and they dissolve when
rinsed with water.
• NEVER IMMERSE THE ELEC-
TRODE UP TO THE CONNEC-
TOR.
TOR.
• ALWAYS KEEP THE CONNEC-
TOR CLEAN AND DRY.
• After use, rinse the electrode
with water to minimize con-
tamination.
tamination.
• Store the electrode
with a few drops of
HI 70300 Storage So-
lution in the protective
cap.
HI 70300 Storage So-
lution in the protective
cap.
• Always replace the protective
cap after use.
DO NOT USE DISTILLED OR
DEIONIZED WATER FOR
STORAGE PURPOSES.
Environment
0 to 50ºC (32 to
122ºF), 95% RH
122ºF), 95% RH
Battery Type
2 x 1.4V
Life
3000 hours
of continuous use
Dimensions
66 x 50 x 25 mm
(2.6 x 2 x 1")
Weight
Skincheck
70 g (2.5 oz.)
Skincheck1
90 g (3.2 oz.)
Initial Preparation
The pH electrode is shipped dry.
Prior to using the
Skincheck
, remove
the protective cap and recondi-
tion the electrode by soaking the
tip (2cm/1") in pH 7.01 buffer so-
lution for a couple of hours.
tion the electrode by soaking the
tip (2cm/1") in pH 7.01 buffer so-
lution for a couple of hours.
Perform the calibration procedure
as described below and the
meter will be ready for taking
measurements.
as described below and the
meter will be ready for taking
measurements.
Specifications
Range
0.00 to 14.00 pH
Resolution
0.01 pH
Accuracy
±0.2 pH
(@20°C/68°F)
Typical EMC
±0.1 pH
Deviation
Calibration
Manual two points
Electrode
combination pH
electrode (included)
Skincheck
HI 1413S/50
Skincheck1
HI 1413B/50
pH
HI 70300
pH
pH
4/10
7
ON
OFF
BACK
TOP VIEW
ON/OFF
SWITCH
BATTERIES
SCREW-TYPE
CONNECTOR
HI 1413S/50
COMBINATION
pH ELECTRODE
PROTECTIVE
CAP
FRONT
LIQUID
CRYSTAL
DISPLAY
BATTERIES
COVER
+ -
pH