Hanna Instruments hi 95759 User Manual

Page of 2
Instruction Manual
HI 95759
Maple Syrup
Transmittance Analyzer
w w w . h a n n a i n s t . c o m
This Instrument is in
Compliance with the CE Directives
Please examine this product carefully. Make sure that the
instrument is not damaged. If any damage occured during
shipment, please notify your Dealer.
Each HI 95759 Transmittance Analyzer is supplied complete
with:
• Six Sample Cuvets
• 9V Battery
• One Light Shield Cap
• Instruction Manual
• Two 5 mL Syringes
• Rigid carrying case
• 30 mL Glycerol (1 bottle)
Note: save all packing material until you are sure that the
instrument works correctly. Any defective item must
be returned in its original packing.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
use of the instrument. Please read it carefully before using
the analyzer. If you need additional technical information,
do not hesitate to e-mail us at tech@hannainst.com or call
1-800-426-6287.
This instrument is in compliance with 
 directives
EN 50081-1 and EN 50082-1.
PRELIMINARY EXAMINATION
GENERAL DESCRIPTION
The  HI 95759 portable microprocessor analyzer measures
the percent light transmittance of maple syrup compared to
analytical reagent grade glycerol. The transmittance value
allows identification of the syrup quality class.
Measurements are made using matched square optical
cuvets having a 10 mm light path.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONS
SPECIFICATIONS
Range
0.0 to 100.0% Transmittance
Resolution
0.1% Transmittance
Precision
±1% @ 75.0% Transmittance
Typical EMC Dev. ±0.1% Transmittance
Light Source
Tungsten lamp with narrow band
interference filter @ 560 nm
Method
Direct Measurement
Light Detector Silicon Photocell
Environment
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type
1 x 9 volt
Auto-Shut off After 10' of non-use
Dimensions
180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight
290 g (10 oz.)
REQUIRED STANDARD
Description
Quantity/test
Glycerol
4 mL
GUIDE TO DISPLAY CODES
SIGNIFICANCE AND USE
The primary characteristic for classifying maple syrup is color.
When syrup is very light in color, as matching the minimum
light transmittance standards, the grade assigned to syrup
is high. When syrup color is dark, the grade is low. Color
classes are expressed in percent of light transmission as
compared to an analytical Glycerol Standard Reference fixed
at one hundred percent transmission. All syrup is considered
to be at minimum density, 66.0 degrees Brix at 68 °F (20
This prompt appears for 1 second
each time the instrument is turned
on.
This indicates that the meter is in a
ready state and calibration can be
performed.
Sampling in Progress. This flashing
prompt appears each time the meter
is performing a measurement.
This indicates that the meter has
been calibrated and measurement
can be performed.
Light over range: the cuvet is not
inserted correctly and an excess
ambient light is reaching the detector.
If the cap is properly installed, then
contact your dealer or the nearest
Hanna Customer Service Center.
The blinking "BAT" indicates that the
battery voltage is getting low and
the battery needs to be replaced.
This indicates that the battery is
dead and must be replaced. Once
this indication is displayed, the meter
will lock up. Change the battery and
restart.
“Configuration”: this indicates that
the meter has lost its configuration.
Contact your dealer or the nearest
Hanna Customer Service Center.
ERROR  MESSAGES
a) on calibration
This indicates that the calibration
procedure failed due to a low signal-to-
noise ratio. In this case press CAL again.
“No Light”: the instrument cannot adjust
the light level. Please check that the
sample does not contain any debris.
WARRANTY
HI 95759 is warranted for two years against defects in workmanship
and materials when used for its intended purpose and maintained
according to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact your dealer. If under warranty, report the
model number, date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be notified
of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service
department and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete
protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments
Inc., Woonsocket, Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
°C), and to have no flavors nor other characteristics extrane-
ous to pure maple syrup.
Color Class
Range (% Transmittance)
Grade A Light Amber
75.0 to 100.0
Grade A Medium Amber
60.5 to 74.9
Grade A Dark Amber
44.0 to 60.4
Grade B for reprocessing
27.0 to 43.9
Substandard
less than 27.0