Hanna Instruments hi 981402 User Manual

Page of 2
Instruction Manual
(HI 981402, HI 981403)
On-line, Waterproof
pH meter with Alarm
http://www.hannainst.com
These Instruments are in
Compliance with the CE Directives
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
WARRANTY
WARRANTY
WARRANTY
WARRANTY
WARRANTY
HI 981402 and HI 981403 are warranted for two years against
defects in workmanship and materials when used for their intended
purpose and maintained according to instructions. The electrode is
warranted for a period of one year. This warranty is limited to
repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-
chase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service de-
partment and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protec-
tion.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instru-
ments Inc., Woonsocket, Rhode Island, 02895 , USA.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the   directives
EN 50081-1 and EN 50082-1.
Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
The meter is supplied with:
• HI 1286 pH electrode for HI981402;
• HI 2114P/2 pH electrode for HI981403;
• Calibration screwdriver;
• 12 VDC power adapter.
Note: Conserve all packing material until the instrument has
been observed to function correctly. Any defective item
must be returned in its original packing.
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
HI 981402 and HI 981403 are pH meters specially designed
to meet the needs of simple continuous monitoring of pH.
The housing has been completely sealed against vapors and
humidity with IP54 rating.
You can simply hang the meter right above the sample to be
tested for continuous measurement.
The HI 1286 and HI 2114P/2 gel-filled pH electrodes are
interchangeable and the BNC connector is protected behind
a waterproof sheath. The unique design of the electrode
provides longer life even in aggressive solutions.
You can even select your own setpoint and be alerted of an
abnormal situation with a flashing LED alarm.
Measurements are highly accurate and the meter can be
calibrated at one or two points.
You no longer need to worry about battery changes either:
the unit runs without interruption on 12 VDC power supply.
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
1. Molded eye
2. Alarm LED
3. pH 4.0 calibration trimmer
4. BNC connector
5. HI 1286 or HI 2114P/2 pH electrode
6. 12 VDC power adapter
7. Power supply connector
8. Protective sheath
9. pH 7.0 calibration trimmer
10. Liquid Crystal Display
pH ELECTRODE CONNECTION & MAINTENANCE
In order to protect the instrument against
vapors and humidity, the BNC connector is
shielded behind a waterproof sheath.
• Slide the protective sheath down. Connect the  pH
electrode to the BNC connector and then slide the
protective sheath back up all the way up to the casing.
Since the protective sheath is rubberized to ensure
maximum waterproof protection, make sure the connec-
tor is completely covered.
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
HI 981402
Electrode
HI 1286 interchangeable pH electrode
Setpoint
3.0 to 11.0 pH
Hysteresis
±0.5 pH around setpoint
Alarm
LED blinks when pH is outside hysteresis range
HI 981403
Electrode
HI 2114P/2 interchangeable pH electrode
Setpoint
6.0 to 9.5 pH
Hysteresis
0.5 pH below setpoint
Alarm
LED starts blinking when pH is over setpoint
COMMON SPECIFICATIONS
Range
0.0 to 14.0 pH
Resolution
0.1 pH
Accuracy (@ 25°C/77°F)
±0.2 pH
Typical EMC Deviation
±0.2 pH
Calibration
Manual with two trimmers
for offset and slope
Casing
IP54
Power supply
External 12 VDC (included)
Dimensions
86 x 94 x 33 mm (3.4 x 3.7 x 1.3”)
Weight
150 g (5.3 oz.)
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE