Hanna Instruments hi 98143-20 User Manual

Page of 2
Instruction Manual
HI 98143
pH/EC isolated
Transmitter
                           0 to 14 pH = 4 to 20 mA
   HI 98143-20
                           0 to 10 mS = 4 to 20 mA
http://www.hannainst.com
These Instruments are in
Compliance with the CE Directives
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
WARRANTY
WARRANTY
WARRANTY
WARRANTY
WARRANTY
HI 98143 is warranted for two years against defects in workman-
ship and materials when used for their intended purpose and maintained
according to instructions. The electrode and the probe are war-
ranted for a period of six months. This warranty is limited to
repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-
chase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service de-
partment and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protec-
tion.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instru-
ments Inc., Woonsocket, Rhode Island, 02895 , USA.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at tech@hannainst.com.
This instrument is in compliance with the   directives.
Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
HI 98143 is a pH/EC transmitter specially designed for long
distance measurements of pH and Electrical Conductivity in
industrial applications. It can be either panel- or wall-
mounted.
The pH and EC signals are isolated to avoid any interference
with the receiving device.
Probes connectors, power supply, contacts and calibration
trimmers are easy to reach directly on the front panel.
For in-line applications the recommended pH electrode is the
HI1001 while the conductivity probe is the HI3001.
HI1001 is a combined double junction pH electrode with
BNC and 3m cable; HI3001 is a 4-ring conductivity probe
with platinum sensors and an internal temperature sensor
to compensate reading for temperature changes.
Both probes have external threads for easy installation in
continuous flow-thru monitoring, for either in-line or tank
applications.
They can resist up to 80°C (176°F) and 6 bar (87 psi).
The instrument is supplied with a differential input. An
HI1283 stainless steel probe can be connected to prevent
potential grounding problems and ensure longer life to the
electrode.
Measurements are accurate and the meter can be cali-
brated at one or two points.
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
  1. pH Offset Calibration Trimmer
2. pH Slope Calibration Trimmer
3. BNC connector for pH Electrode
4. pH electrode (not included)
5. Potential Matching Pin (not included)
6. EC Probe (not included)
7. 12÷24 VDC power adapter (not included)
8. pH Output Terminals
9. EC Output Terminals
10. Power Supply Terminals
11. EC Slope Calibration Trimmer
 12. EC Offset Calibration Trimmer
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range
Range
Range
Range
Range
pH
0.0 - 14.0
EC
0 - 10 mS
Accuracy
Accuracy
Accuracy
Accuracy
Accuracy
pH
±0.5% f.s
(@25°C/77°F)
(@25°C/77°F)
(@25°C/77°F)
(@25°C/77°F)
(@25°C/77°F)
EC
±2% f.s.
Typical
Typical
Typical
Typical
Typical     EMC
EMC
EMC
EMC
EMC
pH
±2% f.s.
Deviation
Deviation
Deviation
Deviation
Deviation
EC
±2% f.s.
Output
Output
Output
Output
Output
pH
4÷20 mA (isolated)
EC
4÷20 mA (isolated)
Temp.Compensation
Temp.Compensation
Temp.Compensation
Temp.Compensation
Temp.Compensation
Automatic from 0 to 60°C with ß=2%/°C
Calibration
Calibration
Calibration
Calibration
Calibration
Manual with two trimmers
pH
Offset: ±2pH; Slope: 80÷110%
EC
Slope: ±20%
pH
pH
pH
pH
pH     Probes
Probes
Probes
Probes
Probes
HI1001 pH electrode (suggested) and
HI1283 Potential Matching Pins (not included)
EC
EC
EC
EC
EC     Probe
Probe
Probe
Probe
Probe
HI3001 (not included) with cell constant 2.1
Power
Power
Power
Power
Power     supply
supply
supply
supply
supply
External, 12 to 24 VDC (not included)
Dimensions
Dimensions
Dimensions
Dimensions
Dimensions 160 x 105 x 31mm (6.2 x 4.1 x 1.2”)
Weight
Weight
Weight
Weight
Weight
280 g (9.9 oz)
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
pH ELECTRODE CONNECTION
• Connect the pH electrode to the BNC connector and the
Potential Matching Pin to the PM terminal.
EC  PROBE CONNECTION
• Probe wires are color coded for an easy operating. Follow
the table below for proper installation.
Terminal #
Probe Wire Color
Cond.Probe 1
green
Cond.Probe 2
white
Cond.Probe 3
red
Cond.Probe 4
blue
Temp.Sens. 1
brown
Temp.Sens. 2
grey
SH
yellow-green
POWER CONNECTION
• Connect a 2-wire cable to the 12÷24 VDC power source
paying attention to its polarity.
OUTPUT CONNECTION
• Connect the pH output terminals to the pH receiver with
a 2-wire cable, and the EC output terminals to the EC
receiver with a second 2-wire cable. Pay attention to the
polarity. At these contacts voltages vary proportionally
to the measured pH and EC values between 4 and 20
mA.
To minimize thermal drifts due to extreme temperature
fluctuations and in case of outdoors measurements, it is
recommended to protect the transmitter with a proper
casing.
• Connect pH electrode, EC probe, power supply and
receiving devices to the transmitter by following the
previous instructions.
• Remove the protective cap from the pH electrode and
immerse it into the sample, at least 4 cm.
• When using the matching pin, dip it together with the
pH electrode; otherwise, connect its terminal to the
reference with a jumper wire.
• Immerse the EC probe in the sample.
• For continuous flow-thru measurements, the probes can
be easily installed in any standard ½” pipe tee. Use
Teflon tape between the probe and the pipe to ensure
a leak free joint. Take care not to overtighten it, as
excessive pressures can cause probe damage.
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
CONNECTIONS
CONNECTIONS
CONNECTIONS
CONNECTIONS
CONNECTIONS