Hanna Instruments ec 214 User Manual

Page of 11
9
8
Power connection
Plug the 12VDC adapter (
HI710005 or HI710006) into the power
supply socket (#10 on page 4; #9 on page 6).
Note: make sure the main line is protected by a fuse.
Probe connection
Connect the conductivity probe to the socket provided (#11 on page
4; #10 on page 6).
Note: Make sure that the instrument has been calibrated before
taking conductivity measurements (see calibration paragraph).
Note: If possible, use plastic beakers to minimize EMC interferences.
TAKING CONDUCTIVITY MEASUREMENTS WITH EC 214
••••• switch the instrument on by pressing the ON/OFF
key.
••••• Immerse the probe in the solution submerging
the holes of the sleeve (0.5 cm below). Tap the
probe lightly on the bottom of the beaker to
remove any air bubbles trapped inside the sleeve.
Take the temperature of the solution with a
CHECKTEMP or a glass thermometer.
••••• Press the “COND/ C” key to select temperature
setting.
••••• Adjust the “TEMPERATURE” knob to display the temperature of the
solution on the LCD.
••••• Press the “COND/ C” key to select conductivity
reading.
••••• Select the appropriate conductivity range.
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
Note: If the display shows "1", there is an over-range condition.
Select the next higher range.
••••• Allow a few minutes for the reading to stabilize. The LCD will
display the temperature compensated conductivity reading.
TAKING CONDUCTIVITY MEASUREMENTS WITH EC 215
••••• switch the instrument on by pressing the ON/OFF
key.
••••• Immerse the probe in the solution submerging
the holes of the sleeve (0.5 cm below). Tap
the probe lightly on the bottom of the beaker
to remove any air bubbles trapped inside the
sleeve.
••••• Adjust the "TEMPERATURE COEFFICIENT" knob to the temperature
coefficient value of the solution (see determination of the tem-
perature coefficient of a solution paragraph).
••••• Select the appropriate conductivity range.
Note: If the display shows "1", there is an over-range condition.
Select the next higher range.
••••• Allow a few minutes for the reading to stabilize. The LCD will
display the temperature compensated conductivity reading.