Hanna Instruments hi 983313-0 User Manual

Page of 2
Instruction Manual
HI 983313-0
HI 983313-1
Panel-Mounted
EC Indicators &
 Controllers
M a n u f a c t u r e r s   s i n c e   1 9 7 8
These Instruments are in
Compliance with the CE Directives
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
WARRANTY
These meters are warranted for two years against defects in work-
manship and materials when used for their intended purpose and
maintained according to instructions. The probe is warranted for a
period of six months. This warranty is limited to repair or replace-
ment free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-
chase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service de-
partment and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protec-
tion.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instru-
ments Inc., Woonsocket, Rhode Island, 02895 , USA.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the 
 directives.
Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
The meter is supplied with:
• Instruction manual;
• Mounting brackets.
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATION
HI 983313-0 and HI 983313-1 are conductivity  indicators
and controllers with a relay output designed for simplicity of
use in a wide range of  applications.
The two models are panel mounted with membrane key-
pads on the front panel and an easy-to-read LCD display.
Both meters automatically compensate for the temperature
variation.
The probe is easy to clean and requires little maintenance.
Measurements are highly accurate and the meters can be
calibrated at one point.
Power supply, wiring and selection are made via the plug-
in terminal blocks on the rear panel.
LED indicators on the front panel identify whether the
controller is in set or measurement mode and if alarm is
active.
GENERAL DESCRIPTION
SPECIFICATIONS
Range
0 to 1999 µS/cm
Resolution
 1 µS/cm
Accuracy (@ 20°C)
   ±2% f.s.
Setpoint
Adjustable through multiturn trimmer
Alarm Condition LED ON and alarm contact closed when
measured EC value is higher than setpoint
Alarm Output
2-contact relay, no fuse protected.
5A, 240 VAC, 30 VDC
Probe
HI 7634-00 EC/TDS probe
Temp.Compensation      Automatic from 5 to 50°C
        (41 to 122°F); ß=2%/ºC
Calibration
Manual, at one point through trimmer
Calibration Solution
HI 7031 (1413 µS/cm)
Power supply:
External
HI 983313-0
 12 VDC
HI 983313-1
  115/230 VAC; 50/60Hz
Dimensions
79 x 49 x 95 mm (3.1 x 1.9 x 3.7”)
Weight:
HI 983313-0
 140 g (4.96 oz)
HI 983313-1
  250 g (8.85 oz)
Recommendations for Users
Before using these products, make sure that they are entirely suitable
for the environment in which they are used. Operation of these instru-
ments in residential areas could cause unacceptable interferences to
radio and TV equipment. The metal band at the end of the probe is
sensitive to electrostatic discharges. Avoid touching this metal band at
all times. During operation, ESD wrist straps should be worn to avoid
possible damage to the probe by electrostatic discharges. Any variation
introduced by the user to the supplied equipment may degrade the
instrument’s EMC performance.
To avoid electrical shock, do not use these instruments when voltage at
the measurement surface exceeds 24 VAC or 60 VDC. Use plastic
beakers to minimize any EMC interferences.
To avoid damage or burns, do not perform any measurement in micro-
wave ovens.
ISTR983313R5  12/01