NI 9233 Operating Instructions and Specifications
14
ni.com
Stopband
The filter significantly attenuates all signals above the stopband 
frequency. The primary goal of the filter is to prevent aliasing. 
Therefore, the stopband frequency scales precisely with the data 
rate. The stopband rejection is the minimum amount of attenuation 
applied by the filter to all signals with frequencies within the 
stopband.
Alias-Free Bandwidth
Any signal that appears in the alias-free bandwidth of the NI 9233 
is not an aliased artifact of signals at a higher frequency. The 
alias-free bandwidth is defined by the ability of the filter to reject 
frequencies above the stopband frequency and it is equal to the data 
rate minus the stopband frequency.
Understanding NI 9233 Data Rates
The frequency of a master timebase (f
M
) controls the data rate (f
s
of the NI 9233. The NI 9233 includes an internal master timebase 
with a frequency of 12.8 MHz, but the module also can accept an 
external master timebase or export its own master timebase. To 
synchronize the data rate of an NI 9233 with other modules that use 
master timebases to control sampling, all of the modules must