National Instruments NI 9233 User Manual

Page of 34
NI 9233 Operating Instructions and Specifications
4
ni.com
Caution
Substitution of components may impair 
suitability for Class I, Division 2.
Caution
For Zone 2 applications, install the system 
in an enclosure rated to at least IP 54 as defined by 
IEC 60529 and EN 60529.
Caution
For Zone 2 applications, connected signals must 
be within the following limit:
Capacitance .......................... 0.2 
μF max
Special Conditions for Hazardous Locations Use in Europe
This equipment has been evaluated as EEx nC IIC T4 equipment 
under DEMKO Certificate No. 03 ATEX 0324020X. Each module 
is marked 
 II 3G and is suitable for use in Zone 2 hazardous 
locations. If you are using the NI 9233 in Gas Group IIC hazardous 
locations or in ambient temperatures of –40 °C
≤ Ta ≤ 70 °C, you 
must use the device in an NI chassis that has been evaluated as 
EEx nC IIC T4, Ex nA IIC T4, or Ex nL IIC T4 equipment.