Fluke 9105 User Manual

Page of 65
48
Hart Scientific
9 Digital Communication Interface
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Set program function to n
pf=n
pf=2
1 to 4
Configuration Menu
Probe Menu
Read R0 calibration parameter
r[0]
r
r0: 999.999
r0: 100.578
Set R0 calibration parameter to n
r[0]=n
r=100.324
98.0 to 104.9
Read ALPHA calibration parameter
al[pha]
al
al: 9.9999999
al: 0.0038573
Set ALPHA calibration parameter to n
al[pha]=n
al=0.0038433
.00370 to .00399
Read DELTA calibration parameter
de[lta]
de
de: 9.99999
de: 1.46126
Set DELTA calibration parameter to n
de[lta]=n
de=1.45
0.0 to 2.9
Read BETA calibration parameter
be[ta]
be
be: 9.999
be: 0.342
Set BETA calibration parameter to n
be[ta]=n
be=0.342
–100.0 to 100.0
Operating Parameters Menu
Set temperature units:
u[nits]=c/f
C or F
Set temperature units to Celsius
u[nits]=c
u=c
Set temperature units to Fahrenheit
u[nits]=f
u=f
Read cutout mode
cm[ode]
cm
cm: {xxxx}
cm: AUTO
Set cutout mode:
cm[ode]=r[eset]/a[uto]
RESET or AUTO
Set cutout to be reset manually
cm[ode]=r[eset]
cm=r
Set cutout to be reset automatically
cm[ode]=a[uto]
cm=a
Read approach setting
ap[proach]
ap
ap:9
ap:5
Set approach setting to degrees
ap[proach]=n
ap=15
0 to 20°C
Read stability
ts
ts
ts:9.9
ts:0.5
Set soak stability to degrees
ts=n
ts=.1
.01 to 4.99°C
Serial Interface Menu
Read serial sample setting
sa[mple]
sa
sa: 9
sa: 1
Set serial sampling setting to seconds
sa[mple]=n
sa=0
0 to 4000
Set serial duplex mode:
du[plex]=f[ull]/h[alf]
FULL or HALF
Set serial duplex mode to full
du[plex]=f[ull]
du=f
Set serial duplex mode to half
du[plex]=h[alf]
du=h
Set serial linefeed mode:
lf[eed]=on/of[f]
ON or OFF
Set serial linefeed mode to on
lf[eed]=on
lf=on
Set serial linefeed mode to off
lf[eed]=of[f]
lf=of
Calibration Menu (WARNING – changing the following calibration values may change the accuracy of the instrument.)
Read B0 calibration parameter
*b0
*b0
b0: 9
b0: 0
Set B0 calibration parameter to n
*b0=n
*b0=0
–999.9 to 999.9
Read BG calibration parameter
*bg
*bg
bg: 999.99
bg: 15625
Set BG calibration parameter to n
*bg=n
*bg=156.25
–999.9 to 999.9
Communications Commands Continued