Fluke 1743 User Manual

Page of 73
1744/1743 
Users Manual 
16 
 
Use only the current probes specified in this manual. If 
you use flexible current probes, wear suitable 
protective gloves or work on de-energized conductors. 
 
Do not expose the Logger to moisture or humidity. 
 
To prevent electrical shock, always connect power 
supply and voltage test leads to the Logger before 
connecting to the load. 
 
All accessories must be approved for 600 V CAT III or 
higher. 
 
Use the Logger only with its original standard 
equipment or with approved optional accessories, as 
listed in Table 2 and Table 3 in this manual. 
 
Connect clip-on current transformers and/or Flexi Set 
to insulated live conductors only. 
 
If measuring sensors are to be connected to non-
insulated live conductors, additional personal 
protective measures must be taken as required by local 
government agencies. 
W
 Caution 
To avoid damage, use the 1744/1743 Power Quality Logger, 
only with the following nominal voltages: 
 
Single-/3-phase, 4-wire (Wye) systems (P-N): 69 V 
to 480 V 
 
3-phase 3-wire (Delta) systems (P-P): 120 V to 830 V 
WX
 Warning 
To avoid electrical shock, or damaging the Logger’s 
internal protective circuitry or weatherproof seal, do not 
open the Logger.