Fluke 1529 Chub-E4 User Manual

Page of 154
the display shows the probe channel or memory location, the probe serial num-
ber, and the conversion type.
Use the
LR buttons to select the channel or memory location. After the
channel is selected, press the Enter button to test the temperature calculation
for the probe. You can enter an input value (
Ω, KΩ, or millivolts) and the corre-
sponding output value (e.g. temperature) is shown. Press the Exit button to exit
to the menu.
7.2.4
Print Probe
The PRINT PROBE function allows you to print probe characteristics to a se-
lected port (either serial or IR). Probe characteristics for a specific channel or
memory location can be printed or for all channel and memory locations. When
45
7 Menu Functions
Probe Menu
1
24.159C
3
24.142C
2
24.634C
4
24.015C
PROBE:
CHAN1
SERIAL:
145-560
TYPE:
CVD
EXIT
ENTER / MENU
UNIT
CONTRAST
1529
CHUB E-4 THERMOMETER READOUT
Figure 16 Test Probe