With this format kohms and millivolts are not converted to ohms and volts.
This command may return the same reading if a new reading has not yet been
obtained. If only new readings are desired, the FORMat:STAMp command can
be used to enable the extended format option and any readings in which the
value of the first field is 0 can be ignored or else the STATus:MEASure? com-
mand can be used to determine if a new reading is available for a desired
channel.
8.4.1.5
FORMat:STAMp?
Returns the extended measurement data format setting. See Section 9.4.1.4,
FETCh? for details.
FORM:STAM?
0
8.4.1.6
FORMat:STAMp <bool>
Enables extended measurement data format with the FETCh?, MEASure? And
READ? commands. On power up this is always set to off.
FORM:STAM ON
8.4.1.7
MEASure? [<chn>]
Returns the most recent measurement for the specified channel.
MEAS? 1
0.0127
This command is the same as the FETCh? [<chn>] command
8.4.1.8
READ? [<chn>]
Returns the most recent measurement for the specified channel.
READ? 2
0.0113
This command is the same as the ?FETCh [<chn>] command.
8.4.1.9
SENSe<chn>:DATA?
Returns the input value for the specified channel.
SENS3:DATA?
100.0291, 0.0000