Fluke 9011 User Manual

Page of 52
S
Access set-point memory
1 100.
Set-point memory 1 location, 100°C currently used
To change to another set-point memory press “UP” or “DOWN”.
4 300.
New set-point memory 4 location, 300°C
Press “SET” to accept the new selection and access the set-point value. Press
“EXIT” to continue and to ignore any changes.
S
Accept selected set-point memory
6.2.2
Set-point Value
The set-point value may be adjusted after selecting the set-point memory and
pressing “SET”.
4 200.
Set-point 4 value in°C
If the set-point value does not need to be changed, press and hold “EXIT” to re-
sume displaying the well temperature. To change the set-point value, press
“SET” and then “UP”.
220.00
New set-point value
When the desired set-point value is reached, press “SET” to accept the new
value and access the temperature scale units selection. If “EXIT” is pressed,
any changes made to the set-point are ignored.
S
Accept new set-point value
6.2.3
Temperature Scale Units
The temperature scale units of the controller maybe set by the user to degrees
Celsius (°C) or Fahrenheit (°F). The units are used in displaying the well tem-
perature, set-point, proportional band, and high limit.
Press “SET” after adjusting the set-point value to change display units.
Un= C
Scale units currently selected
Press “UP” or “DOWN” to change the units.
Un= F
New units selected
Press “SET” to accept the new units or “EXIT” to cancel.
21
6 Controller Operation
Temperature Set-point