Fluke 9011 User Manual

Page of 52
9011 Dual-Well Calibrator
User’s Guide
40
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Display Temperature
Read current set-point
s[etpoint]
s
set: 9999.99 {C or
F}
set: 150.00 C
Set current set-point to n
s[etpoint]=n
s=450
Controller
Range
Read temperature unit
u[nits]
u
u:{C or F}
u:C
Set temperature units:
u[nits]=c/f
C or F
Set temperature units to Celsius
u[nits]=c
u=c
Set temperature units to
Fahrenheit
u[nits]=f
u=f
Read scan function
sc[an]
sc
scan: {ON or OFF} scan: ON
Set scan function:
sc[an]=on/of[f]
ON or OFF
Turn scan function on
sc[an]=on
sc=on
Turn scan function off
sc[an]=of[f]
sc=of
Read scan rate
sr[ate]
sr
srat: 999.9 {C or
F}/min
srat: 10.0
C/min
Set scan rate to
n degrees per
minute
sr[ate]=
n
sr=5
0.1 to 50°C
Read display temperature hold
mode
hm[ode]
hm
hmode: AUTO
hmode: AUTO
Read hold status
ho[ld]
ho
ho:{closed or
open}, 9999.9 {C or
F}
ho: open,
75.0 C
Read temperature
t[emperature]
t
t: 9999.9 {C or F}
t: 55.69 C
Secondary Menu
Read proportional band setting
pr[op-band]
pr
pb: 999.99999
pb: 15.9
Set proportional band to
n
pr[op-band]=n
pr=8.83
Depends on
Configuration
Read cutout setting
c[utout]
c
cu: 9999 {C or F}
cu: 620 C, in
Read heater power
(duty cycle)
po[wer]
po
po: 9999.9
po: 1.0
Ramp and Soak Menu
Read number of programmable
set-points
pn
pn
pn: 9
pn: 2
Set number of programmable
set-points to
n
pn=
n
pn=4
2 to 8
Read programmable set-point
number
n
ps
n
ps3
ps
n: 9999.99 {C or
F}
ps1:
50.00 C
Set programmable set-point num-
ber
n to n
ps
n=n
ps3=50
1 to 8, Instru-
ment Range
Table 2 Controller Communications Commands