3.  USING THE METER 
 
3.1. 
The Electrode  
 
This meter comes equipped with a self-stirring BOD probe, model 
5010, from YSI. Read the accompanying manual prior to installation 
and operation. The 5010 probe need be plugged into the meter only, 
as the meter supplies both its operation power and stirring power. 
 
1.  Prepare the electrode as described in its instruction manual. 
 
2.  Plug the electrode into the DO jack on the back meter panel.  
 
3.  Place the probe into a BOD bottle filled with at least 1 inch of 
water. Allow the probe to warm up for 30 minutes before 
calibration or taking a measurement.