eutech-instruments cyberscan do 1500 do meter User Manual
Instruction Manual
CyberScan DO 1500
3
3. USING THE METER
3.1.
The Electrode
This meter comes equipped with a self-stirring BOD probe, model
5010, from YSI. Read the accompanying manual prior to installation
and operation. The 5010 probe need be plugged into the meter only,
as the meter supplies both its operation power and stirring power.
1. Prepare the electrode as described in its instruction manual.
2. Plug the electrode into the DO jack on the back meter panel.
3. Place the probe into a BOD bottle filled with at least 1 inch of
water. Allow the probe to warm up for 30 minutes before
calibration or taking a measurement.
calibration or taking a measurement.