eutech-instruments conductivity electrode User Manual
Setting the
Standard
in Water
Testing
Standard
in Water
Testing
Built in
temperature
sensor for ATC
temperature
sensor for ATC
Rugged epoxy
housing with
1m cable
housing with
1m cable
68X247782 12/01 Rev 0
OPERATING INSTRUCTIONS
Thank you for selecting an Eutech Instruments
conductivity/TDS cell. Cell constant (K) = 1.0
Dimensions (mm): 132 L x 12 dia.
conductivity/TDS cell. Cell constant (K) = 1.0
Dimensions (mm): 132 L x 12 dia.
BEFORE YOU BEGIN:
Soak electrode in alcohol
for a few minutes to remove oils, if any.
CALIBRATION:
Calibrate according to meter
instructions. ALWAYS use fresh standard:
see selected Eutech calibration standards
below. Select a calibration value equal to
2/3 of the value in the range you are
measuring.
see selected Eutech calibration standards
below. Select a calibration value equal to
2/3 of the value in the range you are
measuring.
DURING MEASUREMENT:
• Avoid particulate matter: allow it to settle
before taking measurements. Don't let
sediment build up on the sensors.
sediment build up on the sensors.
• Avoid air bubble entrapment in the sensor
by gently tapping electrode to dislodge.
• For accurate readings, make sure
conductivity cell sensors are completely
immersed in the sample.
immersed in the sample.
• Rinse with deionized water after every
measurement.
CLEANING:
Soak cell in alcohol for 2 minutes
& then rinse thoroughly in deionized water.
ACCESSORIES:
Conductivity Solutions (480ml per bottle)
EC-CON-1413BT 1413 µS KCI Calibration
EC-CON-1413BT 1413 µS KCI Calibration
Solution
EC-CON-1288BT 12.88 mS KCI Calibration
Solution
EC-442300BT
300 ppm 442 Calibration
Solution
Solution
EC-4423000BT
3000 ppm 442 Calibration
Solution
Solution
Contact your distributor for more solutions.
WARRANTY:
This cell is warrantied for a
period of 6 months from the date of purchase.
If repair, adjustment or replacement is
necessary and has not been the result of
abuse or misuse within the 6 month period,
please return the cell – freight pre-paid –
and correction will be made without charge.
Out of warranty products will be repaired on
a charge basis.
If repair, adjustment or replacement is
necessary and has not been the result of
abuse or misuse within the 6 month period,
please return the cell – freight pre-paid –
and correction will be made without charge.
Out of warranty products will be repaired on
a charge basis.
RETURN OF ITEMS:
Authorization must be
obtained from your distributor before
returning items for any reason.
returning items for any reason.
Note: We reserve the right to make changes, improvements
and modifications to products shown without prior notice.