2.2 Setup
Place the Exer-Rest® device on a flat, stable surface. The minimum clearance
required around the device for safe passage is 130 cm (51 inches) for non-
passage clearance. Lock the device in place by activating the brakes on the
wheels.
A sheet should be placed over the NIMS Memory Foam mattress.
Verify that power settings match local supply voltage. To change the power
setting refer to Chapter 4.6
Plug in the Exer-Rest®.
Special protection comes in the form of an Emergency Stop Button on the right
side of the frame within easy reach of the individual that brings the machine to a
safe stop should discomfort be experienced.
The individual’s weight should not exceed 175 kg or 385 pounds. Individuals
using this device should be cooperative and non-combative.
2.3 Remote Control
The Exer-Rest® device has a wireless touch screen remote to control two modes
of operation: Exercise and Rest.