Insulet Corporation 017 User Manual

Page of 190
137
Cleaning and disinfecting
Keep your PDM test strip port and USB port free of debris and 
liquids. Dirt, dust, blood, and control solution can impair the 
functionality of your PDM or damage it.
If the battery compartment becomes soiled, gently shake out 
the debris or use a dry, lint-free cloth to remove it.
Use the guidelines below to clean or disinfect your PDM:
1. Clean your PDM if it is visibly dirty. Disinfect your PDM at least 
once per week.
2. Be sure to turn off your PDM before cleaning or disinfecting.
3. To clean and disinfect the outside surfaces of the PDM hous-
ing, use Dispatch® Hospital Cleaner Disinfectant Towels with 
Bleach (available online through major retailers).
4. Following the instructions on the label of the Dispatch® Hos-
pital Cleaner Disinfectant Towels with Bleach, wipe the out-
side surfaces of your PDM. Treated surfaces must remain wet 
for a full (2) minutes.
5. Let your PDM air dry completely before checking your BG 
next. We recommend you perform a control solution test to 
ensure the System functions properly. See”Performing a Con-
trol Solution Test
” in Chapter 7, Checking Your Blood Glucose.
6. When finished, thoroughly wash your hands with soap and 
water.
7. If you require assistance, contact call our Customer Care.
Replacing batteries
The PDM uses two (2) AAA alkaline batteries. You can purchase 
them at any grocery, hardware or convenience store.
On average, a pair of batteries powers the PDM for 3 weeks. The 
battery life indicator on the PDM shows the battery power level. 
See ”Personal Diabetes Manager Icons” later in this Appendix for 
the various battery power levels and their respective icons.
Do NOT use IV Prep™ wipes, alcohol wipes, soap, 
detergent, or solvents. Never put any liquid into the 
battery compartment. The PDM is NOT waterproof. 
Do NOT immerse it or place it in or near water. 
While cleaning, do NOT allow debris or liquid to get 
into the test strip port, USB port, battery compart-
ment, or buttons.
AAA alkaline batteries are the preferred battery to 
power the PDM. Do not use any other type of batter-
ies. Never use old or used batteries; the PDM may 
not work properly. Using any kind of batteries other 
than alkaline may reduce battery life and cause dam-
age to the PDM.