Optional Accessories for Use with the Neo Legacy System
CVRx Magnet
The CVRx Magnet can be used to temporarily inhibit the IPG output by placing the solid part of the
magnet over the IPG. The magnet must be kept over the IPG to continue inhibiting the output state.
Once the magnet is removed, the output will continue.
magnet over the IPG. The magnet must be kept over the IPG to continue inhibiting the output state.
Once the magnet is removed, the output will continue.
Figure 4: CVRx Magnet
NOTE: The standard round magnets that are distributed for use with pacemakers and ICDs are readily
available in both cardiology clinics and hospitals. These can also be used to temporarily inhibit the
IPG.
available in both cardiology clinics and hospitals. These can also be used to temporarily inhibit the
IPG.
CVRx Accessory Kit
The CVRx Accessory Kit contains a torque wrench and port plug. Use these to replace those provided
with the IPG in the event that either the torque wrench or port plug is dropped outside the sterile field.
with the IPG in the event that either the torque wrench or port plug is dropped outside the sterile field.
The torque wrench is used to tighten the set screws on the IPG. The port plug can be used to plug a
lead port on the IPG.
lead port on the IPG.
Torque Wrench
Port Plug
Figure 5: CVRx Accessory Kit
CSL Repair Kit Model 5010
The CVRx CSL Repair Kit contains tools and material to repair damage to the insulation and/or
conductor coils of the therapy lead after chronic implantation.
conductor coils of the therapy lead after chronic implantation.