10 Appendix
Patient cables
All visible dirt should be removed using a lint-free cloth moistened with 
a solution of washing up liquid and water. 
For disinfection, clean the patient cables with a cloth moistened with 
70% alcohol solution and then dry.
Deposits of dirt and oxides should be regularly removed from the 
electrode clips by using a toothbrush and a solution of washing-up liquid 
and water.
Important: 
Under no circumstances should the patient cable be immersed in any 
liquid cleaning agent. Nor should they be exposed to steam or hot air 
sterilisation, or chemical sterilisation using ethylene oxide.
NIBP cuffs
Cleaning: wipe the cuff with a cleaning cloth moistened with water and 
ordinary hand detergent. Never use ether or petroleum products. The 
cuff may also be machine-washed. However, hand washing prolongs its 
life.
Disinfection: wipe the cuff with a cleaning cloth moistened with a 70% 
alcohol solution, mild bleaching agent or similar solvent. The cuff may 
be sterilised in an autoclave, or with ethylene oxide.
After washing, allow the cuff to dry, and then replace the rubber bladder.
Important: 
Remove the rubber bladder from the cuff prior to cleaning and 
disinfection. Temperatures over 133 °C (271 °F) damages the cuff, as 
do strong solvents.
SpO
2
Cleaning and disinfection: Saturate a clean, dry gauze pad with the 
cleaning solution. Wipe all surfaces of the sensor and cable with this 
gauze pad.
Saturate another clean, dry gauze pad with sterile or distilled water. 
Wipe all surfaces of the sensor and cable with this gauze pad.
Dry the sensor and cable by wiping all surfaces with a clean, dry gauze 
pad.
Important: 
Caution: Do not sterilise by irradiation, steam, or ethylene oxide.