Ortivus AB 410000409 User Manual

Page of 47
77
Appendix 10
The following ECG cables may be used with MobiMed 300c:
• 10 lead ECG cable.
• 8 lead (Frank vector) ECG cable.
• 5 lead ECG cable.
• 3 lead ECG cable.
10.5.3 SpO
2
 specifications
The pulse oximeter is calibrated to show functional saturation and 
pulse.
The following sensor has been tested and approved:
• Nonin 8000AA Articulated Finger Clip.
10.5.4 NIBP specifications
Oxygen saturation range
70 - 100%
Pulse rate range
18 - 300 bpm
Measurement wavelengths
Red
 660 nm
Infrared
 910 nm
Accuracy SpO
2
Saturation accuracy: 
±
2 digits for 70-
100% (with adult finger clip sensor), 
specified at 
±
1 standard deviation. 
Accuracy is not specified for sensors 
<70% saturation.
Update, saturation and pulse value
Update period 0.3 s
Pulse rate accuracy
±3%, ±1 digit
NIBP measurement range 
60 - 300 mmHg
Ambient pressure
(atmospheric pressure)
355 to 807 mmHg
(-500 to 6000 metres)
NIBP accuracy
±2 mmHg or ±2% depending on 
whichever is the greater
Measurement time
Maximum of 2 minutes. On average, 
less than 40 seconds
Maximum cuff pressure
330 mmHg