Ortivus AB 410000409 User Manual

Page of 47
49
Trends
7
7 Trends
Parameter trends are displayed in Trend view (Figure 31). The Trend 
view is accessed by selecting More:Trends. The Trend view presents 
an immediate overview of the parameters measured during the care 
period. The trended parameters are updated and saved every minute.
The Trends are displayed in two groups each with four parameters. The 
trend curves are named and colour coded on the trend bar. The trend 
bar also indicates the scale values for the displayed parameters. The 
value bar indicates the trend value at marker position.
Figure 31. Trend view
Trend values that fall outside of the parameter limits or scale are 
marked with three asterisks (***) in the trend graph.
The Trend view displays data based on three types of values:
• Single value measurements. The NIBP values that are presented 
consist of a series of single measurements of systolic, diastolic and 
mean blood pressure values.
• Mean value measurements. Pulse, saturation, respiration and heart 
rate values are presented as the mean values of the parameters 
measured during the previous minute.