49
Trends
7
7 Trends
Parameter trends are displayed in Trend view (Figure 31). The Trend
view is accessed by selecting More:Trends. The Trend view presents
an immediate overview of the parameters measured during the care
period. The trended parameters are updated and saved every minute.
The Trends are displayed in two groups each with four parameters. The
The Trends are displayed in two groups each with four parameters. The
trend curves are named and colour coded on the trend bar. The trend
bar also indicates the scale values for the displayed parameters. The
value bar indicates the trend value at marker position.
Figure 31. Trend view
Trend values that fall outside of the parameter limits or scale are
marked with three asterisks (***) in the trend graph.
The Trend view displays data based on three types of values:
• Single value measurements. The NIBP values that are presented
The Trend view displays data based on three types of values:
• Single value measurements. The NIBP values that are presented
consist of a series of single measurements of systolic, diastolic and
mean blood pressure values.
• Mean value measurements. Pulse, saturation, respiration and heart
rate values are presented as the mean values of the parameters
measured during the previous minute.