KROHNE FMCW24G74T User Manual
6
OPERATION
60
OPTIWAVE 7400-24 C
www.krohne.com
12/2015 - 4004901901 - MA OPTIWAVE7400-24 R01 fr
6.3.5 Function description
1.0.0 Information (Info.) menu
1.2.0
Output
1.3.0
History
2.0.0 Supervisor
2.1.0
Quick Setup
2.2.0
Tests
2.3.0
Basic Parameters
2.4.0
Output I
2.5.0
Application
2.6.0
Communication
2.7.0
Display
2.8.0
Conversion
2.9.0
Config/Reset
3.0.0 Service
n/a
Password locked. Menus for factory calibration and qualified
service personnel only.
4.0.0 Master
n/a
Password locked. Menus for factory calibration and qualified
service personnel only.
Menu
No.
No.
Function
Function description
Selection list
Default
1.1.0 IDENT.
1.1.1
SERIAL NUM.
The device serial number.
Read only.
1.1.2
CONV.FIRM.VER
The converter firmware version.
Read only.
1.1.3
SEN.FIRM.VER
The sensor firmware version.
Read only.
1.1.4
HMI.FIRM.VER
The HMI (device display screen)
firmware version.
Read only.
1.1.5
The converter hardware version.
Read only.
1.1.5
The sensor hardware version.
Read only.
1.1.8
The HMI (device display screen)
hardware version.
Read only.
1.1.9
The date of manufacture.
Read only.
1.2.0 OUTPUT
1.2.1
OUTPUT FUNC.
This shows the setting at this time for
the output function (OUTPUT FUNC.).
Read only.
1.2.2
RANGE I
This shows the setting at this time for
the output range (RANGE I).
Read only.
1.2.3
SCALE 4mA
This shows the setting at this time for
the 4 mA setting (SCALE 4mA).
Read only.
1.2.4
SCALE 20mA
This shows the setting at this time for
the 20 mA setting (SCALE 20mA).
Read only.