Data Critical Corporation AVTS1010 User Manual

Page of 55
AlarmView™ System User’s Guide 
 
 
 
{insert part #} Page 13 
 
Section 2: System Specifications 
 
Each hospital or healthcare facility can use up to 6 sets of AlarmView 
Systems, with 24 transmitters each, each set operating on its own 
channel. 
 
 
Transmitter 
 
Parameter 
Setting 
Physical Dimensions 
5.25” x 1.5” x 3” 
(13.3 x 3.8 x 7.6 cm) 
Battery Life 
2 years of normal operation 
Battery Operation 
Approximately 45 minutes 
Output Power 
250 mW nominal 
350 mW maximum 
Frequency Range 
450 MHz to 466 MHz 
Nominal Range 
(no intervening structures) 
Approx. 150 ft (46 m) 
Input Voltage 
+15V, 1.5 amps 
Antenna Terminal Designed 
Impedance 
50 ohms 
Temperature Range 
(normal operation) 
23ºF to 113ºF ( -5ºC to 45ºC) at 5% to 
95% RH, non-condensing 
Temperature Range 
(storage) 
-4ºF to 140ºF ( -20ºC  to 60ºC) at 5% 
to 95% RH, non-condensing 
Altitude 
(normal operation) 
0 ft to 10,000 ft (0 m to 3,048 m) 
Mechanical Shock 
IEC 68-2-27 
Water Resistance 
(transmitter, not power supply) 
IEC 529 Class IPXI  
Radiated 
(radio-frequency electromagnetic 
fields) 
Up to 3V/meter from 80 MHz to 1 
GHz 
Conducted 
(conducted-frequency 
electromagnetic energy) 
Up to 3V/meter from 150 KHz to 80 
MHz 
FCC Compliant 
FCC Part 90