Statistical Analysis Considerations 
The meter calculates statistics from a maximum of 80 readings (For Group 1 through Group 4, a 
maximum of 400 readings can be stored). Note that readings cannot be stored when in DIRECT 
mode. However, statistics on these readings can still be calculated. When the meter is powered 
off or if the working mode is changed (in the programming menu), the DIRECT mode statistics 
will be lost. The following statistical values can be calculated: 
  NO.: Number of readings 
  AVG: Average value 
  Sdev. : Standard deviation (square root of a data set’s variance) 
  MAX: Maximum reading 
  MIN: Minimum reading 
 
Statistical Terms 
Average value (
x
) is the sum of readings divided by the number of readings. 
 
 
 
Standard Deviation (Sdev) 
The sample standard deviation is a statistic that measures how the sample value is 
distributed around the sample mean. The standard deviation of a set of numbers is the root 
mean square of the variance S
 
The variance of a list is the square of the standard deviation of the list, that is, the average of 
the squares of the deviations of the numbers in the list from their mean divided by 
the(number of readings ‐1 ) 
 
Variance:  S
2
=
)
1
/(
)
(
2
n
x
x
  
Standard deviation: S=
2
S
 
 
NOTE: 
Use the DELETE parameter from the programming menu immediately after an outlier or 
erratic reading has been taken. Refer to the Delete function in Programming Menu. 
 
Storage Capacity Overflow 
 
In GROUP mode, if the storage capacity is exceeded, statistics will not be updated, although 
measurements can still be made. If the memory is full, subsequent readings will not be 
included in the statistics. The meter’s display will show “FULL” (in the SINGLE measurement 
mode). 
In DIRECT mode, when the memory is full, the newest reading will replace the oldest reading 
and the statistics will be updated.