Buffalo 09102002-0 External Photos
![Buffalo](https://files.manualsbrain.com/attachments/9b1b53b017a12138833f39c4f8c1f2aa34c2c16e/common/fit/150/50/266e1b8b2619927602ffd298e5482e1aaec72f4e72045ee85919e6306019/brand_logo.png)
REPORT NO: 06J10094-10
DATE: March 2, 2006
FCC ID: FDI-09102002-0
COMPLIANCE CERTIFICATION SERVICES
Page: 29 of 39
This report shall not be reproduced except in full, without the written approval of CCS.
10.2 MEASURMENT UNCERTAINTY 3 GHz – 6 GHz
Ui (1g)
Ui(10g)
Measurement System
Probe Calibration
Probe Calibration
4.80
N
1
1
1
4.80
4.80
Axial Isotropy
4.70
R
1.732
0.707
0.707
1.92
1.92
Hemispherical Isotropy
9.60
R
1.732
0.707
0.707
3.92
3.92
Boundary Effects
1.00
R
1.732
1
1
0.58
0.58
Linearity
4.70
R
1.732
1
1
2.71
2.71
System Detection Limits
1.00
R
1.732
1
1
0.58
0.58
Readout Electronics
1.00
N
1
1
1
1.00
1.00
Response Time
0.80
R
1.732
1
1
0.46
0.46
Integration Time
2.60
R
1.732
1
1
1.50
1.50
RF Ambient Conditions - Noise
3.00
R
1.732
1
1
1.73
1.73
RF Ambient Conditions - Reflections
3.00
R
1.732
1
1
1.73
1.73
Probe Positioner Mechnical Tolerance
0.40
R
1.732
1
1
0.23
0.23
Probe Positioning With Respect to Phantom Shell
2.90
R
1.732
1
1
1.67
1.67
Extrapolation, interpolation, and integration algorithms for
max. SAR evaluation
max. SAR evaluation
3.90
R
1.732
1
1
2.25
2.25
Test sample Related
Test Sample Positioning
Test Sample Positioning
1.10
N
1
1
1
1.10
1.10
Device Holder Uncertainty
3.60
N
1
1
1
3.60
3.60
Power and SAR Drift Measurement
5.00
R
1.732
1
1
2.89
2.89
Phantom and Tissue Parameters
Phantom Uncertainty
Phantom Uncertainty
4.00
R
1.732
1
1
2.31
2.31
Liquid Conductivity - Target
5.00
R
1.732
0.64
0.43
1.85
1.24
Liquid Conductivity - Meas.
8.60
N
1
0.64
0.43
5.50
3.70
Liquid Permittivity - Target
5.00
R
1.732
0.6
0.49
1.73
1.41
Liquid Permittivity - Meas.
3.30
N
1
0.6
0.49
1.98
1.62
Combined Standard Uncertainty
11.66
10.73
Expanded Uncertainty (95% Confidence Interval)
23.32
21.46
Notesfor table
1. Tol. - tolerance in influence quaitity
2. N - Nomal
3. R - Rectangular
4. Div. - Divisor used to obtain standard uncertainty
5. Ci - is te sensitivity coefficient
1. Tol. - tolerance in influence quaitity
2. N - Nomal
3. R - Rectangular
4. Div. - Divisor used to obtain standard uncertainty
5. Ci - is te sensitivity coefficient
RSS
K=2
Std. Unc.(±%)
Uncertainty component
Tol. (±%)
Probe
Dist.
Div.
Ci (1g)
Ci (10g)