Data Critical Corporation 02DR-10100 User Manual

Page of 28
WMTS OPENNET LAN PROGRAMMING AND DIAGNOSTICS
128 
PatientNet Administrator’s Guide, v1.04, 10001003-004, Revision B
All information contained herein is subject to the rights and restrictions on the title page.
10. Click the OP Band button to set the operating band that will be used by the 
RF Module.
Typically, the 608.5-613.0 OP Band is selected. However, where the Patient-
Net System operates in the same hospital as another system (or systems) that 
also use(s) the 608-614 MHz WMTS band, it will be necessary to select one 
of the sub-bands. The sub-band should coordinate with the hospital’s spec-
trum policy.
Additional sub-bands include: 608.5-609.7, 609.6-610.8, 610.7-611.9, and 
611.8-613.0 
WARNING: All transceivers must be programmed to the same OP Band as the Access 
Points. This is an administrative procedure only. Do not attempt to program 
the transceivers with the selected OP Band.
11. Click the Program TX button to program a transceiver and read the pro-
grammed information back for validation. The transceiver is now pro-
grammed.
12. Click the Exit button to return to the Ambulatory Transceiver Programming 
screen.