Data Critical Corporation 02DT-7000 User Manual
TROUBLESHOOTING
218
PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft
All information contained herein is subject to the rights and restrictions on the title page.
Example 12 - False Asystole Alarm
Problem
Small R-Wave Voltage
1. When R-waves are small, the system might not see the signal. A false positive
asystole call and alarm may occur.
Fig. 108. Example 12 - Small R-Wave Voltage
Correction
1. Resolve this by moving the electrodes to a different lead configuration, such
as closer to the heart, to achieve a stronger signal. Increasing the size on the
waveform display will not remedy this situation.
waveform display will not remedy this situation.
Note:
In this situation, the arrhythmia system is using Lead II for analysis and,
although it is of small amplitude, the system is working well.
although it is of small amplitude, the system is working well.
Example 13 - Excessive Pacer Pulses
The strip has a “picket fence” appearance. Note the pulses on the ECG waveform.
Fig. 109. Example 13 - Picket Fencing
This is more easily noted on real-time strips or in the history.
Telectronics manufactures an impedance-based pacemaker that measures the respira-
tion rate and adjusts the pacemaker accordingly. This type of pacemaker emits imped-
ance pulses (18-20 Hz) for adjusting the pacemaker rate based on the patient’s
respiration rate. The front-end device may detect such impedance pulses as pacemaker
tion rate and adjusts the pacemaker accordingly. This type of pacemaker emits imped-
ance pulses (18-20 Hz) for adjusting the pacemaker rate based on the patient’s
respiration rate. The front-end device may detect such impedance pulses as pacemaker