Data Critical Corporation 02DT-7000 User Manual

Page of 62
SYSTEM COMPLIANCE
244 
PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft
All information contained herein is subject to the rights and restrictions on the title page.
DT-4500 Ambulatory Transceiver
UL Classification
The DT-4500 transceiver is classified in accordance with UL 2601-1, Type BF with 
Defibrillator Proof (5000V per AAMI EC-13-1992), and IPX7 (according to IEC 529-
1989). The DT-4500 has been designed to withstand the effects of EMI and meets the 
EMC requirements of EN60601-1-2 (April 1993) and CISPR 11. However, extremely 
high levels of electromagnetic energy (above the levels of EN60601-1-2) may still 
produce interference.
DT-7000/7001 Instrument Transceiver
UL Classification
The DT-7000/7001 transceiver is classified in accordance with UL 2601-1 and IPX1 
(according to IEC 529-1989) on non-patient equipment. The DT-7000/7001 has been 
designed to withstand the effects of EMI and meets the EMC requirements of 
EN60601-1-2 (April 1993) and CISPR 11. However, extremely high levels of electro-
magnetic energy (above the levels of EN60601-1-2) may still produce interference.
DT-4500
WITH RESPECT TO ELECTRIC SHOCK, FIRE,
AND MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH UL2601-1/CAN/CSA
C22.2 NO. 601.1
4P41
DT-700x
WITH RESPECT TO ELECTRIC SHOCK, FIRE,
AND MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH UL2601-1/CAN/CSA
C22.2 NO. 601.1
4P41