Vernier Software & Technology LQ2-LE User Manual

Page of 49
LabQuest App 
 
 
    
15 
Mode: Full Spectrum
 – Full Spectrum 
data-collection mode can only be used 
with Spectrometers and is the default 
mode when a Spectrometer is connected. 
In Full Spectrum mode, Intensity, 
Absorbance, Fluorescence (SpectroVis 
Plus only), or %Transmittance can be 
measured as a function of wavelength. 
 
Detailed instructions for setting up data 
collection with spectrometers are included in each of ou
 
 
Mode: Gas Chromatograph
 – LabQuest 2 supports the use of the Vernier Mini Gas 
Chromatograph (GC). When a Mini GC is plugged in the LabQuest 2 USB port, 
the software will automatically identify the Mini GC and select the Gas 
Chromatograph mode. This mode should not be entered unless a Mini GC is 
attached.  
 
In this mode, various parameters are 
defined for the user to establish a 
temperature and pressure profile 
adequate for the current experiment. 
When the mode is first entered, a set of 
default parameters is displayed. If you 
wish to change these parameters, tap in 
the parameter field to enter a new value 
or adjust the default value by tapping the 
arrow buttons. These values will be 
reflected in the preview of the time-dependent temperature graph displayed at 
right. After setting the parameters, tap OK to initiate the Mini GC warm up. 
 
Note: A new message will appear, “Do not inject until GC is ready.” The 
Mini GC will take a few minutes to warm up and stabilize. When the Mini GC is 
ready for injection, the message will read, “Inject and select Collect 
simultaneously.”
  
 
For more detailed information on the Vernier Mini Gas Chromatograph, see 
 
Mode: Drop Counting
 – In this mode, you can choose to calibrate drops so that the 
volume of titrant is recorded in units of milliliters. Choose Calibrate from the 
Sensors menu. The Vernier Drop Counter is set to use a default calibration of 28 
drops/mL. If you choose to calibrate the sensor, the volume of an individual drop 
will be determined by the number of drops that pass through the Vernier Drop 
Counter and dividing by the total volume of the drops. Follow the instructions on 
the screen to complete a custom calibration. Alternately, the Equation tab allows 
you to input a previously determined value for drops/mL.  Enter the drops/mL 
value (e.g., 28.0), then choose OK.