ZM-521PA/531PA
Measuring Parameters
Measuring waveforms:
ECG, respiration in impedance method, pulse
Measuring numeric data:
Heart rate, respiration rate, SpO
2
, pulse rate
Transmitting Data
Waveform data:
ECG, respiration, pulse wave
Numeric data:
SpO
2
, pulse rate
Status information:
Battery replacement, battery level*, alarm suspended,
pause monitoring*, patient confirmed*, ECG lead, pacing
detection, electrode detachment, electrode impedance*,
ECG off*, respiration method (impedance)*, SpO
pause monitoring*, patient confirmed*, ECG lead, pacing
detection, electrode detachment, electrode impedance*,
ECG off*, respiration method (impedance)*, SpO
2
status,
respiration status*, channel ID, time constant (3.2 s), type
of transmitter, transmitter code number*, transmitter serial
number*
of transmitter, transmitter code number*, transmitter serial
number*
* The items marked with * are transmitted only when the protocol is “51”.
Display
Display size:
2.2 inch TFT color LCD
Viewing area:
44.16 (H) × 33.12 (V) mm
Resolution:
320 (H) × 240 (V) dots
Displayed Data
Numeric and waveform screen:
ECG (one waveform from lead I, II, III, Va or Vb), heart
rate, pulse rate, respiration rate, SpO
rate, pulse rate, respiration rate, SpO
2
, message, battery
level, QRS/pulse sync mark, pulse bar graph, ECG lead
Waveform review screen:
ECG or pulse wave of past 10 minutes
Numeric review screen:
Heart rate or pulse rate, respiration rate and SpO
2
at 1
minute interval for past 10 minutes
Check electrodes screen:
ECG for checking electrode attachment
ECG
ECG measurement
Channels: 4
Input dynamic range:
Channels: 4
Input dynamic range:
±10 mV or more
Electrode offset potential tolerance: ±500 mV or more
Input impedance:
Input impedance:
5 MΩ or more
Common mode rejection ratio:
95 dB or more
IEC 60601-2-27: 2005 50.102.10 complied
Pacing pulse detection:
amplitude ±2 to 700 mV, duration 0.1 to 2 ms
IEC 60601-2-27: 2005 complied
Based upon pacemaker pulse rejection capability