Nihon Kohden Corporation ZM-531PA User Manual
Operator’s Manual ZM-520PA/521PA/530PA/531PA
63
ECG and Respiration Monitoring
When the electrodes are attached and the ECG leads are connected, the heart rate, ECG
waveform, respiration rate and respiration waveform appear on the receiving monitor. Refer to the
operator’s manual of the receiving monitor for details.
waveform, respiration rate and respiration waveform appear on the receiving monitor. Refer to the
operator’s manual of the receiving monitor for details.
When 6 leads are used on this transmitter, up to 8 leads (I, II, III, aVR, aVL, aVF, Va and Vb) of
ECG waveforms can be displayed on the receiving monitor. The heart rate is also measured.
When 3 leads are used, one channel ECG waveform of lead II can be displayed on the receiving
monitor. Refer to the operator’s manual of the monitor for details.
ECG waveforms can be displayed on the receiving monitor. The heart rate is also measured.
When 3 leads are used, one channel ECG waveform of lead II can be displayed on the receiving
monitor. Refer to the operator’s manual of the monitor for details.
Pacing marks
White pacing marks
are displayed at
the points where
a pacing spike is
detected.
White pacing marks
are displayed at
the points where
a pacing spike is
detected.
Heart rate
QRS sync mark
Respiration rate
ECG lead
You can change the
ECG lead with the
You can change the
ECG lead with the
▼
and
▼ keys.
ECG waveform
Filter: on
Sweep speed:
Sweep speed:
12.5 mm/s
Waveform sensitivity: auto
CAUTION
Turn off the power of mobile phones, small wireless devices and other devices which
produce strong electromagnetic interference around a patient (except for devices
allowed by the hospital administrator). Radio waves from devices such as mobile
phones or small wireless devices may be mistaken as pulse waves or respiration
waves and the displayed data may be incorrect.
produce strong electromagnetic interference around a patient (except for devices
allowed by the hospital administrator). Radio waves from devices such as mobile
phones or small wireless devices may be mistaken as pulse waves or respiration
waves and the displayed data may be incorrect.
NOTE
• ECG cannot be monitored on a neonate using this transmitter.
• When performing defibrillation, set the hum filter to ON on the receiving monitor. The
• When performing defibrillation, set the hum filter to ON on the receiving monitor. The
waveform recovery may become slow due to electrode polarization when the hum filter
is set to OFF.
is set to OFF.