Instructions for Use | Device Setup 
 
 
 
Connecting a Circuit 
34 
 
Active PAP Circuits 
 
Bacteria filter 
Proximal pressure port  
Active exhalation valve line connection 
Tubing 
 
1.  Connect the bacteria filter (1) on the circuit to the Inspiratory Port. 
2.  Connect the proximal pressure line to the Proximal Pressure Port (2). 
3.  Connect the active exhalation valve pressure line to the active exhalation valve line connection (3).   
Active Flow Circuits 
 
Bacteria filter 
Proximal pressure port  
Active exhalation valve line connection 
Flow sensor cable connector 
Tubing 
 
1.  Connect the bacteria filter (1) on the circuit to the Inspiratory Port. 
2.  Connect the proximal pressure line to the Proximal Pressure Port (2). 
3.  Connect the active exhalation valve pressure line to the active exhalation valve line connection (3).