Respironics Inc. 1127941 User Manual

Page of 100
 
 
Instructions for Use | Alarms and System Messages 
 
 
 
Alarms and System Messages 
59 
 
System Messages 
Message 
Cause 
External DC Source Disconnected 
Power is supplied from an external DC source and that power source is 
disconnected or depleted. 
Depleted External DC 
The external DC power source is depleted. 
Check External DC 
An external battery is connected but cannot supply sufficient power. 
Check for a faulty cable, bad connection, or bad battery. 
Depleted Detachable Battery 
The detachable battery is depleted. 
Internal Battery Not Charging – Temperature 
The system is unable to charge the internal battery due to temperature. 
Internal Battery Not Discharging – Temperature 
The internal battery is unable to power the device due to high or low 
temperature. Change the environmental temperature or relocate the 
device. 
Detachable Battery Not Charging – Temperature 
The system is unable to charge the detachable battery due to 
temperature. 
Detachable Battery Not Discharging – Temperature 
The system is unable to charge the detachable battery due to 
temperature. 
Start On Battery 
Trilogy Evo Universal is turned on and AC power is not detected. 
Reconnect AC power. 
Power is restored after losing all power and AC power is not detected. 
Reconnect AC power. 
Circuit Mismatch 
ActivePAP, passive, or dual limb/active flow circuits 
only 
The circuit type selected for use does not match the circuit connected to 
Trilogy Evo Universal for 3 consecutive breaths. Check the circuit type. 
Connect a circuit that matches the prescription. 
Unsupported Accessory Connected 
An unsupported accessory is connected to Trilogy Evo Universal. 
FiO
2
 sensor not calibrated 
When FiO
2
 monitoring is enabled and an FiO
2
 sensor is connected but was 
not calibrated   
Replace FiO
2
 sensor 
The FiO
2
 sensor is at the end of life or is no longer functioning. 
Pulse Oximeter Connected 
Indicates a pulse oximeter is now connected. 
Mainstream Capnography Sensor Connected 
Indicates a Mainstream Capnography Sensor is now connected