Respironics Inc. 1127941 User Manual

Page of 100
 
 
Instructions for Use | Trilogy Evo Universal Accessories 
 
 
 
Oxygen 
82 
 
Dual Limb Active Exhalation Valve 
The Philips Respironics Dual Limb Active Exhalation Valve is a dual limb circuit accessory. The valve will close during inspiration and open during 
exhalation to divert flow into the expiratory limb for venting outside of the breathing circuit. This device is intended for multi-patient use. Clean 
and disinfect or clean and sterilize this item between uses on different patients. 
For help setting up, using, or maintaining this device, contact Philips Respironics.   
Warnings: 
• 
To reduce the likelihood of disconnection, and to prevent degraded or adverse ventilator performance, use this accessory only with a 
compatible ventilator.  This device is compatible and validated for use with Philips Respironics Trilogy Evo Universal-series ventilators. 
Before use, ensure the compatibility of the ventilator and all of the parts, materials, and accessories you plan to use to connect to the 
patient.  
• 
Do not modify this equipment.  
• 
This product may become contaminated by bodily fluids and expired gases. Dispose according to 
your local regulations. 
 
Storage and Handling: Avoid direct sunlight. 
Disposal: This product may become contaminated by bodily fluids and expired gases. Dispose according to 
your local regulations. 
Instructions for Use: Install according to the illustration. For cleaning and disinfection instructions, see the 
“Cleaning and Disinfection” chapter. 
Oxygen 
Adding Oxygen to the Device 
Delivered oxygen concentration varies with changes in flow in the circuit. The following may have an impact on oxygen concentration:  
• 
Pressure settings 
• 
Patient Tidal Volume 
• 
Peak Inspiratory Flow 
• 
I:E Ratio 
• 
Respiratory rate 
• 
Circuit leak rate 
• 
Oxygen flow rate 
 
Low Flow Oxygen Warnings 
• 
When administering fixed-flow supplemental oxygen, the oxygen concentration may not be constant. The inspired oxygen concentration 
will vary, depending on the pressures, patient flow, and circuit leak. Substantial leaks may reduce the inspired oxygen concentration to 
less than the expected value. Use appropriate patient monitoring, such as pulse oximeter with alarm, as medically indicated. 
• 
This device DOES NOT alarm for loss of the low flow oxygen supply. 
• 
Do not connect the device to an unregulated or high-pressure oxygen source. 
• 
The device may result in incorrect flow and tidal volume measurements and improper operation of related alarms if you add low flow 
oxygen directly into the patient circuit or mask instead of directly adding it into the oxygen inlet on the back of the ventilator. 
• 
Turn off oxygen when the device is not in use. When the device is not in operation and the oxygen flow remains on, oxygen delivered into 
the tubing may accumulate within the device’s enclosure. 
• 
Do not operate the ventilator in the presence of flammable gasses. This could cause a fire or explosion. 
 
To add oxygen to the circuit, the oxygen supply must comply with the local regulations for medical oxygen. The oxygen 
flow into the oxygen valve cannot exceed 30 l/min and the pressure cannot exceed 10 psi. For instructions, see the “Device 
Setup” chapter.