ODSONIC INTERNATIONAL CO. LIMITED 9100N User Manual

Page of 4
【Quick Installation Guide─Infant Temperature Monitor 
v1.0
】 
                                                                                                                                                          3   
 
                                                                                                        07-2017 / v1.0 
 
 
 
 
 
 
 
                                                             
▲Sound Level Settings Cycle
 (Pic.6)
 
 
  2.2.2    Units of Temperature Options 
(℃ / ℉)
 
❶ The 
℃ icon will start flashing, indicating this option is ready for setting. 
❷ By pressing the 
 Button, it will cycle between the 
℃ / ℉ options
 
❸ To confirm the Unit of Temperature 
(℃ / ℉)
,Press the 
 Button, and it will automatically initiate 
the next Option. 
 
2.2.3    Body Temperature High/Low Threshold Options
 
❶ The 
H     
icon
 
and the Value will start flashing, indicating this option is ready for setting. 
❷ Press the 
 Button, it will display the selectable values for High-Temperature Threshold. 
(Table 4)
 
Press the 
⊝ 
button to cycle through the different values. 
✦Reminder~ You must set the High-Temperature Threshold Value first. To set the Low-Temperature 
Threshold Value, follow step ❷, refer to 
(Table 5) 
for the Low-Temp. Values. 
❸ After confirming the High/Low Threshold Values, press the 
Button, and it will automatically 
initiate the next Option. 
2.2.4    Ambient Temperature High/Low Threshold Options
 
❶ The 
H     
and the Value will start flashing, indicating this option is ready for setting. 
❷ Press the 
 Button, it will display the selectable values for High-Temperature Threshold.
 (Table 6)
 
Press the 
⊝ 
button to cycle through the different values. 
✦Reminder~ You must set the High-Temperature Threshold Value first. To set the Low-Temperature 
Threshold Value, follow step ❷, refer to 
(Table 7) 
for the Low-Temp. Values. 
 
❸ After confirming the High/Low Threshold Values, press the 
Button, and it will return to the 
Sound Level Options   
 
3. 
FAQ
 
(1) Is the temperature accurate immediately after powering on the device? 
* Please wait 30 seconds after the Baby Temperature Monitor is powered on, to allow the sensor to 
warm up and display a more accurate reading.   
 
(2)
 
What is the effective detection range of the Baby temperature Monitor? How does it determine 
where to take the temperature? 
* The effective range is a 35 degrees cone from the center of the Thermal Image Receiver, the sensor 
will automatically search for the point with the highest-temperature (from the forehead) and 
continously monitor the temperature.   
 
(3) How do I save the Options and Settings after setting them?
 
* After finishing setting all Options above, please remember to press the ⦿ Button to return to the 
main menu and to save all Options and Settings! 
 
(4) How do I clear the Options? / How do I unpair the Sensor and the Receiver?
 
*If you wish to clear the options and unpair the sensor/reciever, simply hold down the recessed Reset 
Button next to to the Power Button using a small and slim object. (a paper clip.) 
 
(5) What happened to the display? Why did it turn off while I was changing the options? 
* The Liquid Crystal Display (Receiver) will automatically hibernate after five seconds of inactivity. Press 
the ⊕ Button to continue choosing the options.   
 
Icon Flashing: 
(Waiting for Confirmation)