LNL-1 manual Fujtecom., Inc. 2014
25
7-5-3
“Date/time Setting Screen”
This screen is used for setting the date/time of the analyzer main body. Use the up/down arrow keys to change
the year/month/day/time.
Use the right/left keys to move to other setting items and press the ENTER key to save the setting.
7-5-4
“Measurement Time Setting Screen”
This screen is used to set the logger measurement starting time and the measurement interval. (Measurement
time: 2 hours, measurement interval: 1 to 60 seconds, and maximum sampling numbers: 7200 data/day)
Use the up/down arrow keys to change the measurement starting time and the measurement interval.
Use the right/left keys to move to other setting items and press the ENTER key to save the setting.
When the measurement starting time is decided, the measurement ending time is automatically set to be two
hours later.
Caution
If sufficient collection time is not kept, the measurement time returns to the previous measurement time
when the ENTER key is pressed to save the setting. Be careful so that the measurement time and the
collection time do not overlap each other. If the measurement time and the collection time must overlap each
other, set the collection time to be 4 hours or more.
Measurement time
Measurement interval