7-5-10 “Data Delete Selection Screen”
The loggers that can be connected via wireless communication
can be searched/displayed.
Logger communication exclusion mark
This mark is set when you do not wish to overwrite the setting.
Select the logger ID and use the DELETE key to perform
setting/release.
The exclusion mark is also attached in the following cases.
* When the logger power supply is turned on/off
* This mark is set to detect the logger in the collection time period.
* Automatically attached in the communication time period, but not
attached immediately after power supply is turned on
* 10 minutes before logging
* Automatically attached also during logging time
When the ENTER key is pressed, the data delete confirmation screen is displayed.
Delete execution icon
Delete cancellation icon
Use the right/left arrow keys to select the icon.
When the delete execution icon is selected and the ENTER key is pressed, delete is executed.
When the delete cancel icon is selected and the ENTER key is pressed, delete is cancelled and the screen returns
to the “Data Delete Selection Screen”.
Caution
If delete is executed, the data saved in the logger and analyzer SD card are deleted.
Data clear function via analyzer (three functions)
Data within analyzer
Data within logger
Logger ID editing
Deleted
Not deleted
Logger setting
Not deleted
Deleted
Data delete
Deleted
Deleted