45
Alarm level
Condition
Message
Cause
Alert
Stuck Key
! STUCK KEY
Key pressed for more than 3 minutes.
ESC,OK to clear
ESC,OK to clear
Key pressed for more than 3
minutes.
minutes.
Alert
End of
Suspend
Suspend
! END OF SUSPEND
Stopped at []. ESC,OK to clear
Pump suspended manually or
automatically.
automatically.
Alert
Low
Reservoir
Reservoir
! LOW RESERVOIR
[] U or less left [] hours remaining,
replace patch soon. ESC,OK to clear
replace patch soon. ESC,OK to clear
Low insulin level reached.
Alert
Auto Off
Advisory
Advisory
! AUTO OFF ADVISORY
Escalates to Auto-Off alarm if not
cleared within 15 minutes. ESC,OK to
clear
cleared within 15 minutes. ESC,OK to
clear
Alert indicating that no Patch
status received during the set
time limit.
status received during the set
time limit.
Alert
Patch
Expiration
Advisory
Expiration
Advisory
! PATCH EXPIRATION ADVISORY
Patch expires after [] hours, change
patch soon. ESC,OK to clear
patch soon. ESC,OK to clear
Patch expires in less than the
set time limit.
set time limit.
Alert
Patch
Expiration in
1 Hour
Expiration in
1 Hour
! PATCH EXPIRATION IN 1 HOUR
Patch expires in less than 1 hour,
change patch now. ESC,OK to clear
change patch now. ESC,OK to clear
Patch expires in less than 1
hour.
hour.
Alert
Low PDM
! LOW PDM
Low PDM. ESC,OK to clear
The battery is low.
Alert
RF Out of
Range
Range
! RF OUT OF RANGE
Move devices closer and wait 1 minute.
ESC,OK to clear
ESC,OK to clear
Pump Base is out of range.