Olympus Winter & Ibe GmbH ESG200 User Manual

Page of 110
22
Product description and functional principles
For selected CELON bipolar applicators the electrosurgical generator also provides:
 - Radiofrequency induced thermotherapy (RFITT)
 - Resistance Controlled Automatic Power (RCAP) for RFITT
 - Automatic instrument recognition
 - Automatic wear-out detection
The functional principles of theses features are described in the sections below.
3.3.1 CQM – Contact Quality Monitoring for the neutral electrode
When using split type neutral electrodes for monopolar electrosurgery, the CELON ELITE 
ESG-200 is able to detect unintended detachment of the neutral electrode from the patient. 
The indicator for the contact quality monitor (CQM) shines green while the contact between 
the split type neutral electrode and the skin of the patient is within an acceptable resistance 
range. If the contact between the split type neutral electrode and the patient’s skin is 
insufficient, then an alarm tone sounds, a warning message is displayed and the contact 
quality monitor indicator shines red.
Using non-split type neutral electrodes is not as safe as using split-type ones as CQM is not 
able to detect any detachment of the neutral electrode from the patient.
For detailed information on the safe and correct use of neutral electrodes refer to the chapter 
“Neutral electrode operation” on page 53.
3.3.2 HPCS – High Power Cut Support
All cutting modes are supported by HPCS. This feature optimizes the start of the cutting 
procedure by applying high power to the tissue to support immediate spark ignition and to 
reduce the risk of mechanical cutting.
3.3.3 Autostart for the bipolar mode SoftCoag
If the impedance is within a defined range, then Autostart feature permits automatic 
activation of the output power as soon as both electrodes of the electrosurgical instrument 
touch the tissue. It is possible to set a time delay so that the automatic activation does not 
start immediately when the electrodes are in contact with the tissue but only after a certain 
time. Consequently, activating by foot switch is not necessary.
Autostart is available only for the bipolar mode SoftCoag.
3.3.4 FSM – Fast Spark Monitor
This feature ensures smooth and reproducible cutting although the tissue characteristic are 
varying, e.g. in muscle and fat.
3.4 Features for selected CELON bipolar applicators
The following features are compatible only with selected CELON bipolar applicators, e.g. 
CelonProSleep plus or CelonProBreath.
DRAFT 
– 
2017-01-17