Olympus Winter & Ibe GmbH ESG200 User Manual

Page of 110
97
Appendix A
APPENDIX A
Alarm system
An alarm condition is indicated about one second after detection and securing.
Log file for alarm conditions
The alarm system provides a log of occurrences of alarm conditions. The eldest 
log is overwritten by the latest log when the capacity of 4098 logs is reached. Only 
authorized service personnel can read log data.
If the alarm system is powered down, then the log is still maintained by an internal 
electrical power source. In case of a total loss of power, i.e. mains electricity as well 
as internal power source, the complete log is erased immediately.
Alarm tones
Depending on the priority of an error, the alarm tones sound different.
High priority errors
E0004, E0008...E0011, E0013, E0015, E0018, E0024...E0037, E0039...E0102, 
E0116...E0119, E0122...E0133, E0137, E0142...E0178, E0197, E0198, E0203...
E0211, E0213, E0216...E0412, E0415...E0437, E0439, E0440, E0443...E0455, 
E0458...E0484, E0492...E0605, E0620...E0659, E0662...E0664, E0672...E0680, 
E0682, E0684...E0686, E0691, E0697
For these error codes the following values apply:
Duty cycle ....................................................Intermitting signal (number of burst = 10)
Frequency ................................................................................................. 200/300 Hz
Volume ...................................................................................................... ≥ 65 dB (A)
Medium priority errors
E0012, E0016, E0017, E0019...E0022, E0038, E0104, E0105, E0108, E0109, 
E0120, E0121, E0134, E0136, E0179...E0186, E0188...E0196, E0199...E0201, 
E0212, E0215, E0413, E0414, E0438, E0457, E0488, E0489, E0665, E0681
For these errors codes the following values apply:
Duty cycle ......................................................Intermitting signal (number of burst = 3)
Frequency ................................................................................................. 200/300 Hz
Volume ...................................................................................................... ≥ 65 dB (A)
Low priority errors
E0001, E0002,  E0006, E0115, E0140, E0141, E0187, E0202, E0214, E0441, 
E0619, E0660, E0683, E0687...E0690, E0698
For these errors codes the following values apply:
Duty cycle ......................................................Intermitting signal (number of burst = 2)
Frequency ................................................................................................. 200/300 Hz
Volume ...................................................................................................... ≥ 65 dB (A)
DRAFT 
– 
2017-01-17