Operator Manual – ORP System 
 
 
Revision: 05 
 
 
Effective Date: 17-Dec-15 
 
 
Document Number: YST-UDoc-OPM-05 
 
 
 
109 of 116 
Table 6 – Table 7 – IEC 60601-1-2:2007 Table 2 Requirements 
The Prodigy Sleep Monitor is intended for use in the electromagnetic environment specified below. The 
customer or the user of the Prodigy Sleep Monitor should assure that it is used in such an environment. 
Immunity Test 
IEC 60601 Test Level 
Compliance Level 
Electromagnetic 
environment – Guidance 
Electrostatic 
Discharge (ESD) 
IEC 61000-4-2 
±6 kV Contact 
±8 kV Air 
±6 kV Contact 
±8 kV Air 
Floors should be wood, 
concrete or ceramic tile. 
If floors are covered with 
synthetic material, the 
relative humidity should 
be at least 30%. 
Electrical Fast 
Transient/Burst 
IEC 61000-4-4 
±2 kV for power supply 
lines 
±2 kV for power supply 
lines 
Mains power quality 
should be that of a typical 
commercial or hospital 
environment. 
Surge 
IEC 61000-4-5 
±1 kV line(s) to line(s) 
±1 kV line(s) to line(s) 
Mains power quality 
should be that of a typical 
commercial or hospital 
environment. 
Voltage Dips, 
short interruptions 
and voltage 
variations on 
power supply 
input lines 
IEC 61000-4-11 
<5% UT 
(>95 % dip in UT) 
for 0,5 cycle 
40% UT 
(60% dip in UT) 
for 5 cycles 
70% UT 
(30% dip in UT) 
for 25 cycles 
<5% UT 
(>95% dip in UT) 
for 5 sec 
<5% UT 
(>95 % dip in UT) 
for 0,5 cycle 
40% UT 
(60% dip in UT) 
for 5 cycles 
70% UT 
(30% dip in UT) 
for 25 cycles 
<5% UT 
(>95% dip in UT) 
for 5 sec 
Mains power quality 
should be that of a typical 
commercial or hospital 
environment. If the user 
of the Prodigy Sleep 
Monitor requires 
continued operation 
during power mains 
interruptions, it is 
recommended that the 
Prodigy Sleep Monitor 
be powered from an 
uninterruptible power 
supply or a battery. 
Power Frequency 
Magnetic Field 
(50/60 Hz) 
IEC 61000-4-8 
3 A/m 
3 A/m 
Power frequency 
magnetic fields should be 
at levels characteristic of 
a typical location in a 
typical commercial or 
hospital environment. 
NOTE UT is the a.c. mains voltage prior to application of the test level.